Skip to main content
. 2022 May 10;28(1):77–95. doi: 10.1007/s10741-022-10235-9

Table 1.

Overview of most common CMR sequences

Sequence characteristics Applications Limitations
Native imaging (without contrast injection)
Cine imaging

Balanced-SSFP

(segmented, ECG-gated, multiple cardiac phases)

Possible 3D acquisition (but lower spatial/temporal resolution, longer acquisition times)

LV/RV volumes, systo/diastolic function, wall thickness, LV/RV mass

Banding (“India Ink”) artifacts highlight fat/water boundaries (e.g. fat infiltration)

Susceptibility to magnetic field inhomogeneities (e.g. metal implants, poor shimming)

Acquired over multiple heartbeats (limited by irregular RR-intervals or breathing movements)

Lower temporal resolution than echocardiography

Spoiled-GRE

(segmented, ECG-gated, multiple cardiac phases)

Used in case of metal implants (lower susceptibility to metal artifacts)

Lower contrast (blood-to-myocardium) resolution

Limited by arrhythmias/

breathing movements (similarly to SSFP cine)

Real-time GRE or SSFP (single-shot, ungated, multiple cardiac phases) Used to track beat-to-beat cardiac motion (e.g. septal movements in suspected tamponade/ constrictive physiology; diaphragmatic movements in suspected paralysis) Low spatial and temporal resolution
Black-blood imaging T1- or PD- or T2-weighted double-IR FSE (segmented or single-shot, ECG-gated, triggered to a single diastolic cardiac phase)

LV/RV morphology and tissue-characterization (e.g. fatty infiltration)

T2-weighted fat-saturated IR-FSE used as an alternative to T2-weighted STIR sequences for oedema detection

Still fluids (subendocardial bloodpool, effusions…) appear hyperintense
STIR T2-weighted triple-IR FSE (segmented or single-shot, ECG-gated, triggered to a single diastolic cardiac phase)

Intra/extracellular oedema, such as in inflammation and acute necrosis (qualitative/ semiquantitative detection of hyperintense areas)

Markedly hypointense areas correspond to myocardial haemorrage or calcifications

Quantification of oedema is time consuming

Cardiac segments close to the surface coil may appear hyperintense

Still fluids (subendocardial bloodpool, effusions…) appear hyperintense

T1-mapping

MOLLI (8–11 single-shot, ECG-gated IR-SSFP images, all acquired at the same systolic or diastolic cardiac phase with different TIs)

Other IR- or SR- sequences are possible alternatives

Native T1 (quantitative): increased by inflammation, oedema, vasodilation, fibrosis, amyloid; decreased by fat, iron

Limited spatial resolution

Needs motion correction algorithms (image misalignment may cause incorrect T1 calculation)

T2-mapping MESE (Multi echo spin echo), GraSE (Gradient echo spin echo) or T2-prepared bSSFP: 3–4 images, all acquired at the same systolic or diastolic cardiac phase with different T2-weighing Native T2 (quantitative); increased by inflammation, oedema; decreased by iron

Limited spatial resolution

Needs motion correction algorithms (image misalignment may cause incorrect T1 calculation)

T2*-mapping GRE multiecho: 6–8 segmented, ECG-gated images, all acquired at the same systolic or diastolic cardiac phase with different T2*-weighing Native T2* (quantitative); decreased by iron deposition (haemochromatosis, haemorrage)

Limited spatial resolution

Susceptibility to magnetic field inhomogeneities (e.g. metal implants, poor shimming)

Phase contrast

Spoiled-GRE

(segmented, ECG-gated, multiple cardiac phases)

Possible 3D/4D acquisition (but longer acquisition times and motion artifacts)

Flow quantification (quantitative), across cardiac valves, aortic or pulmonary vessels

Limited spatial and temporal resolution compared to Doppler-echocardiography

Unsuitable for vessels as small as the coronary arteries

Inaccurate in case of magnetic field inhomogeneities

Post-contrast imaging (after Gd-based contrast injection)
Perfusion IR- or SR-, GRE or SSFP during Gd-based contrast injection

Myocardial perfusion (qualitative/semiquantitative)

Quantitative myocardial perfusion with specific dual-bolus or dual-sequence techniques

Limited spatial resolution

Possible dark rim artifact in the subendocardial blood-to-myocardium interface

Early enhancement (EGE) IR GRE or IR-SSFP (segmented or single-shot, ECG-gated, triggered to a single systolic/diastolic cardiac phase), with a TI set to null the thrombus Intracardiac thrombus detection Selection of a wrong nulling time makes EGE image inaccurate
Late enhancement (LGE)

IR GRE or IR-SSFP (segmented or single-shot, ECG-gated, triggered to a single systolic/diastolic cardiac phase), with a TI set to null the normal myocardium

Possible 3D acquisition (but longer acquisition times and more motion artifacts)

Extracellular Gd deposition (qualitative/semiquantitative, increased by necrosis, fibrosis, amyloid deposition but also intense extracellular oedema) with excellent contrast-to-noise ratio

Markedly hypointense areas within LGE correspond to no-reflow areas

Quantification of fibrosis is time consuming

Detection of diffuse fibrosis remains challenging

Selection of a wrong nulling time makes LGE image inaccurate; PSIR (phase sensitive inversion recovery) LGE less dependent on TI

ECV-mapping Same as T1 mapping (MOLLI or other sequences) Extracellular Gd deposition (qualitative): increased by necrosis, amyloidosis, amyloid, but also extracellular oedema

Limited spatial resolution

Needs a pre- and post-contrast acquisition, with perfect image fusion

Needs blood haematocrit for ECV calculation

Vascular imaging
CEMRA

3D GRE during Gd-based contrast injection

Possible time-resolved CEMRA acquisition (but lower spatial resolution)

Aorta and its branches, pulmonary arteries and its branches,

Needs contrast injection

ECG-ungated (unsuitable for coronary arteries)

Lower spatial resolution than CT

3D-whole heart

3D balanced-SSFP

(segmented, ECG-gated, respiratory navigator- gated, triggered to a single cardiac phase)

Coronary artery anatomy

Cardiac arterial and venous connection anatomy

Long acquisition time

Limited by arrhythmias/

breathing movements

Common CMR sequences are based on an FSE, spoiled-GRE or a SSFP structure (readout), with variable T1/PD/T2 weighing depending on the chosen parameters (flip angle, repetition time, echo time), sometimes preceded by an IR- or SR- prepulse (to selectively invert or saturate specific tissues)

CEMRA, contrast-enhanced magnetic resonance angiography; CT, computed tomography; ECV, extracellular volume; FSE, fast spin-echo; GRE, gradient echo; IR, inversion recovery; LV, left ventricle; MOLLI, modified Look-Locker inversion recovery; PD, proton density; RV, right ventricle; SR, saturation recovery; (b) SSFP, (balanced) steady-state free-precession; STIR, short-tau inversion recovery; TI, inversion time