TABLE 2.
Parameter | Entity/actor | Specification |
---|---|---|
PET scanner calibration | Technologist | Must perform daily/weekly/monthly scanner quality assurance and vendor-recommended maintenance procedures (e.g., replace weak transmission sources for dedicated PET scanner); must ensure that output values are acceptable and manually entered on form/electronic database |
PET scanner calibration constancy check | Technologist | Must perform constancy phantom (e.g., 68Ge cylinder) scan (preferably NIST-traceable or equivalent to gather information on uniformity as well) at least weekly and after each calibration |
Radionuclide calibrator | Physicist | Must be calibrated to 18F using NIST-traceable source or equivalent either by site or by calibrator manufacturer |
NIST = National Institute of Standards and Technology.
Only shaded rows are mandatory for profile conformance; white row is recommended and may be mandatory in future profile updates.