Site administrator |
3.6.1.1: site accreditation/qualification maintenance |
Same |
Site administrator |
3.6.2: imaging facility personnel |
Same |
Medical physicist |
3.6.3: amyloid PET acquisition scanner |
Same |
Medical physicist |
3.6.3.1.1: radionuclide calibrator |
Same |
Medical physicist |
3.6.3.1.2: scales and stadiometers |
Not required |
Medical physicist |
3.6.3.1.4: clocks and timing devices |
Not required |
Medical physicist |
3.6.4.1: uniformity and calibration |
Not required |
Medical physicist |
3.6.4.2: resolution |
Same |
Medical physicist |
3.6.4.3: noise |
Same |
Medical physicist |
3.6.4.4: amyloid PET specific phantom measurements |
Same |
Medical physicist |
4.1: performance assessment: image acquisition site |
Same |
Technologist |
3.1.3.1.2: radiopharmaceutical activity calculation or schedule |
Dose structured report not required |
Technologist |
3.1.3.1.3: radiopharmaceutical administration route |
Excludes saline flush and checking for infiltration |
Technologist |
3.2.1.1: timing of image data acquisition |
Same |
Technologist |
3.2.1.2: subject positioning |
Does not cover strict serial scan positioning |
Technologist |
3.2.1.3: scanning coverage and direction |
Same |
Technologist |
3.2.1.4: scanner acquisition mode parameters: PET acquisition |
Same except does not cover if scan is stopped and restarted |
Technologist |
3.2.1.4: scanner acquisition mode parameters: CT acquisition |
Not required |
Technologist |
3.3.1: imaging data reconstruction |
Same except point spread function is allowed |
Image Analyst |
4.4: performance assessment: image analysis workstation |
Not required |