Differences Between Profile Specifications and ADNI 2 Protocol Specifications

Actor Profile section for reference ADNI 2 (58)
Site administrator 3.6.1.1: site accreditation/qualification maintenance Same
Site administrator 3.6.2: imaging facility personnel Same
Medical physicist 3.6.3: amyloid PET acquisition scanner Same
Medical physicist 3.6.3.1.1: radionuclide calibrator Same
Medical physicist 3.6.3.1.2: scales and stadiometers Not required
Medical physicist 3.6.3.1.4: clocks and timing devices Not required
Medical physicist 3.6.4.1: uniformity and calibration Not required
Medical physicist 3.6.4.2: resolution Same
Medical physicist 3.6.4.3: noise Same
Medical physicist 3.6.4.4: amyloid PET specific phantom measurements Same
Medical physicist 4.1: performance assessment: image acquisition site Same
Technologist 3.1.3.1.2: radiopharmaceutical activity calculation or schedule Dose structured report not required
Technologist 3.1.3.1.3: radiopharmaceutical administration route Excludes saline flush and checking for infiltration
Technologist 3.2.1.1: timing of image data acquisition Same
Technologist 3.2.1.2: subject positioning Does not cover strict serial scan positioning
Technologist 3.2.1.3: scanning coverage and direction Same
Technologist 3.2.1.4: scanner acquisition mode parameters: PET acquisition Same except does not cover if scan is stopped and restarted
Technologist 3.2.1.4: scanner acquisition mode parameters: CT acquisition Not required
Technologist 3.3.1: imaging data reconstruction Same except point spread function is allowed
Image Analyst 4.4: performance assessment: image analysis workstation Not required