TABLE VI. Measurement of SpO2 and HR With Setting of 10% Pulse Amplitude, Small Finger, and 2.5% Respiration Artifact.
| HR Setting | 30 bpm | 60 bpm | 120 bpm | 180 bpm | ||||
|---|---|---|---|---|---|---|---|---|
SpO
|
SpO
|
HR | SpO
|
HR | SpO
|
HR | SpO
|
HR |
| Setting | (%) | (bpm) | (%) | (bpm) | (%) | (bpm) | (%) | (bpm) |
| 99 | 99 | 30 | 99 | 60 | 99 | 120 | 99 | 181.08 |
| 98 | 98 | 30 | 98 | 60 | 98 | 120 | 98 | 181.35 |
| 97 | 97 | 30 | 97 | 60 | 97 | 120 | 97 | 181.16 |
| 95 | 95 | 30 | 95 | 60 | 95 | 120 | 95 | 181.11 |
| 90 | 90 | 30 | 90 | 60 | 90 | 120 | 90 | 180.57 |
| 85 | 85 | 30 | 85 | 60 | 85 | 120 | 85 | 181.44 |
| 80 | 80 | 30 | 80 | 60 | 80 | 120 | 80 | 180.53 |
| 70 | 70 | 30 | 70 | 60 | 70 | 120 | 70 | 181.66 |