Table 3. Procedural characteristics and outcomes.

BAV patients (n=109) TAV patients (n=137) p-value
Valve type Self-expanding valve 91 (83.5%) 113 (82.5%) 0.143
Balloon-expandable valve 5 (4.6%) 14 (10.2%)
Mechanically expandable valve 13 (11.9%) 10 (7.3%)
Transfemoral access 105 (96.3%) 133 (97.1%) 0.735
Predilatation 106 (97.2%) 131 (95.6%) 0.739
Post-dilatation 50 (45.9%) 56 (40.9%) 0.432
Oversizing by annulus perimeter, % 6.52 (1.05-13.33) 9.72 (4.31-15.13) 0.014
Implantation depth*, mm 4.5 (1.7-8.0) 6.0 (3.9-11.0) <0.001
Conversion to SAVR 4 (3.7%) 1 (0.7%) 0.174
Cardiopulmonary resuscitation 5 (4.6%) 3 (2.2%) 0.472
Coronary obstruction 1 (0.9%) 1 (0.7%) 1.000
Myocardial infarction 1 (0.9%) 0 (0%) 0.443
Bleeding 10 (9.2%) 10 (7.3%) 0.593
Aortic dissection 5 (4.6%) 0 (0%) 0.016
Stroke 1 (0.9%) 1 (0.7%) 1.000
Tamponade 2 (1.8%) 1 (0.7%) 0.586
Acute kidney injury 6 (5.5%) 5 (3.6%) 0.697
PPMI 10 (9.2%) 27 (19.7%) 0.022
New atrial fibrillation 3 (2.8%) 8 (5.8%) 0.394
Blood transfusion 14 (12.8%) 22 (16.1%) 0.479
Severe patient-prosthesis mismatch 3 (2.9%) 4 (3.0%) 1.000
Echocardiography before discharge
LVEF, % 59.7 (52.0-64.8) 58.8 (52.1-64.2) 0.563
Max velocity, m/s 2.44±0.51 2.40±0.46 0.531
Mean gradient, mmHg 12.89±5.34 12.35±5.01 0.423
Aortic valve area, cm2 1.57±0.29 1.61±0.33 0.332
Moderate/severe PVL 7 (6.5%) 16 (11.9%) 0.162
Data are presented as n (%) or mean±SD or median (interquartile range, IQR). p-values in bold are statistically significant. *Implantation depth was measured by fluoroscopy on the non-coronary cusp direction. Life-threatening bleeding. BAV: bicuspid aortic valve; LVEF: left ventricular ejection fraction; PPMI: permanent pacemaker implantation; PVL: paravalvular leakage; SAVR: surgical aortic valve replacement; TAV: tricuspid aortic valve