Table 2.
Uncertainty analysis of standard solutions for IMS and Nicolet Evolution 300 in first and second series of measurements.
| Standard Solution (µg Fe2+/L) | Standard Sample Uncertainty (µg Fe2+/L) |
First Series | Second Series | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| IMS (µg Fe2+/L) | IMS Prec. (µg Fe2+/L) | IMS Uncert. (µg Fe2+/L) | Nicolet (µg Fe2+/L) | Nicolet Prec. (µg Fe2+/L) |
Nicolet Uncert. (µg Fe2+/L) |
IMS (µg Fe2+/L) |
IMS Prec. (µg Fe2+/L) | IMS Uncert. (µg Fe2+/L) |
Nicolet (µg Fe2+/L) | Nicolet Prec. (µg Fe2+/L) |
Nicolet Uncert. (µg Fe2+/L) |
||
| 25 | ±0.25 | - | - | - | - | 28.7 | ±88.3 | ±9.4 | 28.2 | ±795.2 | ±28.2 | ||
| 50 | ±0.50 | 54.8 | ±27.0 | ±5.2 | 54.7 | ±275.5 | ±16.6 | 44.4 | ±22.0 | ±4.7 | 41.3 | ±795.2 | ±28.2 |
| 75 | ±0.75 | - | - | - | - | - | - | 75.6 | ±36.0 | ±6.0 | 79.4 | ±795.2 | ±28.2 |
| 100 | ±1.00 | 98.4 | ±25.0 | ±5.0 | 98.1 | ±275.5 | ±16.6 | 97.0 | ±21.1 | ±4.6 | 98.6 | ±795.2 | ±28.2 |
| 125 | ±1.25 | - | - | - | - | - | - | 122.6 | ±49.0 | ±7.0 | 124.1 | ±795.2 | ±28.2 |
| 150 | ±1.50 | - | - | - | - | - | - | 152.2 | ±36.0 | ±6.0 | 149.7 | ±795.2 | ±28.2 |
| 200 | ±2.00 | 196.6 | ±18.4 | ±4.3 | 196,0 | ±275.5 | ±16.6 | 201.5 | ±28.0 | ±5.3 | 200.9 | ±795.2 | ±28.2 |
| 500 | ±5.00 | 513.8 | ±28.0 | ±5.3 | 511.2 | ±275.5 | ±16.6 | - | - | - | - | - | - |
| 1000 | ±10.00 | 993.3 | ±27.0 | ±5.2 | 994.9 | ±275.5 | ±16.6 | - | - | - | - | - | - |