Skip to main content
. 2023 Jan 17;23(3):1058. doi: 10.3390/s23031058

Table 2.

Uncertainty analysis of standard solutions for IMS and Nicolet Evolution 300 in first and second series of measurements.

Standard Solution (µg Fe2+/L) Standard Sample Uncertainty
(µg Fe2+/L)
First Series Second Series
IMS (µg Fe2+/L) IMS Prec. (µg Fe2+/L) IMS Uncert. (µg Fe2+/L) Nicolet (µg Fe2+/L) Nicolet Prec.
(µg Fe2+/L)
Nicolet Uncert.
(µg Fe2+/L)
IMS
(µg Fe2+/L)
IMS Prec. (µg Fe2+/L) IMS Uncert.
(µg Fe2+/L)
Nicolet (µg Fe2+/L) Nicolet Prec.
(µg Fe2+/L)
Nicolet Uncert.
(µg Fe2+/L)
25 ±0.25 - - - - 28.7 ±88.3 ±9.4 28.2 ±795.2 ±28.2
50 ±0.50 54.8 ±27.0 ±5.2 54.7 ±275.5 ±16.6 44.4 ±22.0 ±4.7 41.3 ±795.2 ±28.2
75 ±0.75 - - - - - - 75.6 ±36.0 ±6.0 79.4 ±795.2 ±28.2
100 ±1.00 98.4 ±25.0 ±5.0 98.1 ±275.5 ±16.6 97.0 ±21.1 ±4.6 98.6 ±795.2 ±28.2
125 ±1.25 - - - - - - 122.6 ±49.0 ±7.0 124.1 ±795.2 ±28.2
150 ±1.50 - - - - - - 152.2 ±36.0 ±6.0 149.7 ±795.2 ±28.2
200 ±2.00 196.6 ±18.4 ±4.3 196,0 ±275.5 ±16.6 201.5 ±28.0 ±5.3 200.9 ±795.2 ±28.2
500 ±5.00 513.8 ±28.0 ±5.3 511.2 ±275.5 ±16.6 - - - - - -
1000 ±10.00 993.3 ±27.0 ±5.2 994.9 ±275.5 ±16.6 - - - - - -