Skip to main content
. 2023 Jan 12;47(1):27–41. doi: 10.4093/dmj.2022.0271

Table 1.

Comparison of continuous glucose monitoring systems

Dexcom G6 Dexcom G7 Guardian Sensor 3 Guardian Sensor 4 FreeStyle Libre FreeStyle Libre 2 FreeStyle Libre 3 Eversense Eversense E3
Company Dexcom Dexcom Medtronic Medtronic Abbott Abbott Abbott Senseonics Senseonics
Size, mm 45×30×15 27.3×24×4.6 35×28×9.5 35×28×9.5 35×5 35×5 21×2.9 18.3×3a 18.3×3a
FDA approved year 2018 Awaiting 2018 Awaiting 2017 2020 2022 2018 2022
Approved age ≥2 years ≥2 years ≥2 years ≥14 years ≥18 years ≥4 years ≥4 years ≥18 years ≥18 years
Sensor type Real-time Real-time Real-time Real-time Intermittent scanning Intermittent scanning Real-time Real-time (implantable) Real-time (implantable)
Wear time 10 days 10.5 days 7 days 7 days 14 days 14 days 14 days 90 days 180 days
Warm-up period 2 hours 27 minutes 2 hours 2 hours 1 hour 1 hour 1 hour 24 hours 24 hours
Calibration required No No Yes (2 times/day) No No No No Yes (2 times/day) Yes (2 times/day for 21 days, 1 time/day after day 21)
Alarms for hypo/hyperglycemia Yes Yes (can be delayed up to 6 hours) Yes Yes No Yes (optional) Yes (optional) Yes (on-body vibration) Yes (on-body vibration)
Integrity with insulin pumps Yes Yes Yes Yes No Yes Yes No No
Platform Dexcom clarity Dexcom clarity CareLink Personal CareLink Personal Libreview Libreview Libreview
LibreLinkUp Eversense DMS Eversense DMS
Accuracy (MARD) 10.0% 8.2% (arm), 9.1% (abdomen) 8.7% (arm), 9.6% (abdomen) 10.6% 11.4% (arm) 9.2% (adults), 9.7% (pediatrics) 9.2% (adults), 9.7% (pediatrics) 8.8% 9.1%
Accuracy in low glucose range (MAD) 10.9 mg/dL (glucose level <54 mg/dL) 8.5 mg/dL (arm), 10.3 mg/dL (abdomen) (glucose level 40–60 mg/dL) NA NA 11.3 mg/dL (glucose level <100 mg/dL) 9.1 mg/dL (adults), 8.8 mg/dL (pediatrics) 9.1 mg/dL (adults), 8.8 mg/dL (pediatrics) NA 7.5 mg/dL (glucose level ≤60 mg/dL)
Bluetooth free range 6 m 6 m 6 m 6 m NA NA 10 m 7.6 m 7.6 m

FDA, U.S. Food and Drug Administration; MARD, mean absolute relative difference; MAD, mean absolute difference; NA, not applicable.

a

Implantable sensor size.