Comparison of present and future insulin pumps integrated with CGM

Stages Tubed
Tubeless
SAP+PLGS HCLS AHCLS HCLS or AHCLS SAP HCLS HCLS
Model DIA:CONN G8 MiniMed 670G/770G MiniMed 780G T:Slim X2 Eopatch M Eopatch X Omnipod 5
Manufacturer G2E Medtronic Medtronic Tandem Eoflow Eoflow Insulet
Size, cm 9.5×5.4×2.4 9.6×5.36×2.44 9.68×5.36×2.49 7.95×5.08×1.52 4.95×3.9×1.45 4.95×3.9×1.45 5.2×3.9×1.45
Weight, g 84 (without battery and empty reservoir) 106 (without battery and empty reservoir) 106 (without battery and empty reservoir) 112 (with battery and full reservoir) 29.4 (without battery and empty reservoir) 29.4 (without battery and empty reservoir) 26 (without battery and empty reservoir)
Basal rate, units/hr 0.02–30 0.025–35 0.025–35 0.1–15 0.05–15 0.05–15 0.05–30
Basal rate increments, units/hr 0.01 <1 u: 0.025 <1 u: 0.025 0.001 0.05 0.05 0.05
>1 u: 0.05 >1 u: 0.05
>10 u: 0.1 >10 u: 0.1
Bolus, units 0.05–30 0.025–25 0.025–25 0.05–25 0.05–25 0.05–25 0.05–30
Reservoir, units 300 300 300 300 200 200 200
Durable, day NA NA NA NA 3.5 3.5 3
Integration with CGM Dexcom G6 Metronic 3 Sensor Guardian 4 sensor Dexcom G6 Dexcom G6 Dexcom G6 Dexcom G6
Software Diaconn Care Web CareLink CareLink t:connect web Eobridge Eobridge N/A
Mobile App Diaconn App MiniMed Mobile MiniMed Mobile t;connect mobile Narsha Narsha Omnipod 5
Calibration No Yes (2 times/day) No (with Guardian 4 sensor) No No No No
Algorithm location NA NA NA NA Mobile App Mobile App In the Pod

CGM, continuous glucose monitoring; SAP, sensor-augmented pump; PLGS, predicted low glucose suspend; HCLS, hybrid closed-loop system; AHCLS, advanced hybrid-closed loop system; NA, not applicable.