TABLE 1.

Device performance using respiratory samples compared to real-time reverse transcriptase PCR results

Test type No. of samplesa
Sensitivity (95% CI) (%)b Specificity (95% CI) (%)
TP FN FP TN
iHealth 87 33 0 32 72.5 (63.6–80.3) 100.0 (89.1–100.0)
On/Go 90 30 0 32 75.0 (66.3–82.5) 100.0 (89.1–100.0)
BinaxNOW 84 36 0 32 70.0 (61.0–78.0) 100.0 (89.1–100.0)
QuickVue 98 22 4 28 81.7 (73.6–88.1) 87.5 (71.0–96.5)
InBios 79 41 0 32 65.8 (56.6–74.2) 100.0 (89.1–100.0)
a

TP, true positives; FN, false negatives; FP, false positives; TN, true negatives.

b

Sensitivity (TP/TP+FN) and specificity (TN/TN+FP) were calculated. The upper and lower bounds for 95% confidence intervals (CIs) are shown.