Skip to main content
. 2022 Dec 15;11(1):e02286-22. doi: 10.1128/spectrum.02286-22

TABLE 2.

Device sensitivity using Omicron and non-Omicron variant respiratory samples for three CT value groups

Test type CT value Data for:
Omicron samples
Non-Omicron samples
No. of samples:
Sensitivity (95% CI) (%) No. of samples:
Sensitivity (95% CI) (%)
Tested RCAT positive Tested RCAT positive
iHealth Total 89 65 73.0 (62.6–81.9) 31 22 71.0 (52.0–85.8)
<20 27 27 100.0 (87.2–100.0) 10 9 90.0 (55.5–99.80)
20–25 36 34 94.4 (81.3–99.3) 12 12 100.0 (73.5–100.0)
>25 26 4 15.4 (4.4–34.9) 9 1 11.1 (0.3–48.3)
On/Go Total 89 68 76.4 (66.2–84.8) 31 22 71.0 (52.0–85.8)
<20 27 27 100.0 (87.2–100.0) 10 9 90.0 (55.5–99.80
20–25 36 36 100.0 (90.3–100.0) 12 12 100.0 (73.5–100.0)
>25 26 5 19.2 (6.6–39.4) 9 1 11.1 (0.3–48.3)
BinaxNOW Total 89 62 69.7 (59.0–79.0) 31 22 71.0 (52.0–85.8)
<20 27 27 100.0 (87.2–100.0) 10 9 90.0 (55.5–99.80
20–25 36 33 91.7 (77.5–98.3) 12 11 91.7 (61.5–99.8)
>25 26 2 7.7 (1.0–25.1) 9 2 22.2 (2.8–60.0)
QuickVue Total 89 75 84.3 (75.0–91.1) 31 23 74.2 (55.4–88.1)
<20 27 27 100.0 (87.2–100.0) 10 9 90.0 (55.5–99.8)
20–25 36 36 100.0 (90.3–100.0) 12 12 100.0 (73.5–100.0)
>25 26 12 46.2 (26.6–66.6) 9 2 22.2 (2.8–60.0
InBios Total 89 59 66.3 (55.5–76.0) 31 20 64.5 (45.4–80.8)
<20 27 27 100.0 (87.2–100.0) 10 9 90.0 (55.5–99.8)
20–25 36 31 86.1 (70.5–95.3) 12 11 91.7 (61.5–99.8)
>25 26 1 3.9 (0.1–19.6) 9 0 0.0 (0.0–33.6)