TABLE 5.

Performance characteristics of the BLESSED protocol for the detection of ESBL-PE in clinical settings

Diagnostic performance characteristic Median value (95% CI) compared with:
Rectal swab culture Postenrichment broth culture
Sensitivity (%) 96 (78–100) 90 (74–98)
Specificity (%) 93 (87–97) 98 (93–100)
Positive predictive value (%) 71 (57–82) 87 (68–95)
Negative predictive value (%) 99 (95–100) 98 (95–99)
Positive likelihood ratio 14 (7.4–27) 37.3 (12.1–114.8)
Negative likelihood ratio 0.05 (0.01–0.32) 0.10 (0.03–0.29)