Skip to main content
. 2023 Jan 27;415(7):1357–1369. doi: 10.1007/s00216-023-04527-8

Table 3.

Lower limit of detection (LLOD), lower limits of quantification (LLOQ), and linear range of exogenous standards, using a maximum of ten measuring points between 1 fg/mL and 10 ng/mL

Vit-D metabolite LLOD (pg/mL) LLOQ (pg/mL) Linear range (pg/mL)
1,25(OH)2D3 1 12.5 1–1000
1,25(OH)2D2 1 12.5 10–1000
24,25(OH)2D3 1 12.5 10–1000
24,25(OH)2D2 1 12.5 10–1000
25(OH)D3 1 12.5 10–1000
25(OH)D2 1 12.5 10–1000