Table 1.

Comparison of the Overall Quantitative Fit Factors or Simulation Workplace Protection Factors (SWPFs) Obtained by Participants; and the Proportion of Participants Achieving an Overall Fit Factor or SWPF >100 a

Power Fit Factor or SWPF Before Chest Compression During Chest Compression After Chest Compression P
Value
Off Overall score, median (IQR) [range] 1,869 (617–4333)
[265–7,407]
1,748 (378–6,881)
[84–12,170]
1,243 (669–3,881)
[130–9,187]
.69
Score >100, % 100 87.5 100 .37
On Overall score, median (IQR) [range] 3,576 (2,128–6,109)
[428–8,522]
4290 (2,048–4,931) [502–6,381] 4,135 (2,913–6,890)
[496–7,720]
.09
Score >100, % 100 100 100
a

Before, during and after chest compression, while wearing the HALO device with the power on and power off.