Table 6.

Dose rates for each tomosynthesis mammography projection in comparison with European guidelines acceptable dose reference values

Breast thickness AGD mean AGD mean DBT-SCV European guidelines ESD mean ESD mean DBT-SCV
Right Left Acceptable dose Right Left
CC MLO CC MLO CC MLO CC MLO
2–3 0.54 0.71 0.68 0.57 0.73  < 1.5 2.64 2.61 2.74 3.28 3.4
3.1–4 0.97 1.01 0.88 0.98 0.94  < 2.0 3.4 4.99 3.28 4.73 3.72
4.1–5 1.19 1.21 1.23 1.18 1.22  < 2.5 3.66 5.03 4.03 4.74 4.54
5.1–6 1.42 1.45 1.39 1.31 1.34  < 3 4.33 4.34 4.34 3.95 4.57
6.1–7 1.65 1.74 1.68 1.72 1.52  < 4.5 5.06 5.26 4.99 4.73 5.02
7.1–8 2.29 2.68 2.88 2.94 2.04  < 6.5 4.66 5.44 5.21 4.98 5.66