Insertion depth of intraocular lens injector systems into a clear corneal incision.

Injector system Manufacturer Insertion depth (mm)
Mean SD 95% CI
AuotonoMe Alcon Inc 2.75 0.18 2.69–2.81
Ultrasert Alcon Inc 2.62 0.19 2.54–2.70
MultiSert Hoya Medical Singapore Ltd 5.53 0.37 5.41–5.65
Kowa original injector Kowa Company Ltd 6.04 0.33 5.93–6.15
Accuject 1.6-1P Medicel AG 4.54 0.29 4.38–4.70
Accuject 1.8-1P Medicel AG 6.00 0.55 5.83–6.17
Accuject 2.2-1P Medicel AG 6.06 0.43 5.92–6.20
Ergoject 2.2-TL Medicel AG 5.50 0.45 5.25–5.74
Viscoject-Bio 2.2 Medicel AG 5.57 0.50 5.34–5.80
Ray Sert PLUS Rayner Intraocular Lenses Ltd 6.33 0.63 6.17–6.48
RayOne Rayner Intraocular Lenses Ltd 6.30 0.64 6.05–6.55
Bluemixs 180 Carl Zeiss Meditec AG 6.20 0.51 6.04–6.37
SkyJet Carl Zeiss Meditec AG 6.64 1.23 5.96–7.33