Table 1. HILIC Method Validation Results for Imprecision, LOQ, Linearity, and Carryover.
analyte | average RSD (% CV) | LOQ (μM) | linearity (R2) | carryover (%) |
---|---|---|---|---|
CRN | 8.9 | 11.2 | 0.996 | 0.0014 |
Val | 15 | 115 | 0.993 | 0.059 |
t-Hcy | 20 | 6.85 | 0.985 | 0.025 |
GUAC | 6.1 | 0.456 | 0.996 | –0.056 |
Leu | 5.5 | 24.2 | 0.997 | 0.061 |
CRE | 6.1 | 115 | 0.989 | 0.014 |
MA | 7.2 | 4.29 | 0.997 | 0.16 |
MMA | 8.6 | 1.51 | 0.997 | 0.47 |
EMA | 6.1 | 0.867 | 0.997 | 0.032 |
3-HGA | 10 | 0.753 | 0.998 | 0.50 |
2-HGA | 5.4 | 9.65 | 0.993 | 0.12 |
OA | 5.6 | 0.297 | 0.998 | 0.026 |
MCA | 5.7 | 0.305 | 0.998 | 0.029 |
LPC 20:0 | 8.2 | 0.0809 | 0.998 | 0.032 |
LPC 22:0 | 7.5 | 0.0159 | 0.998 | 0.00018 |
LPC 24:0 | 9.3 | 0.0294 | 0.991 | 0.0051 |
LPC 26:0 | 14 | 0.0264 | 0.992 | 0.010 |
Ile | 11 | 9.85 | 0.994 | 0.050 |
aIle | 11 | 0.347 | 0.992 | 0.065 |