Skip to main content
. 2023 Feb 1;95(6):3187–3194. doi: 10.1021/acs.analchem.2c03098

Table 1. HILIC Method Validation Results for Imprecision, LOQ, Linearity, and Carryover.

analyte average RSD (% CV) LOQ (μM) linearity (R2) carryover (%)
CRN 8.9 11.2 0.996 0.0014
Val 15 115 0.993 0.059
t-Hcy 20 6.85 0.985 0.025
GUAC 6.1 0.456 0.996 –0.056
Leu 5.5 24.2 0.997 0.061
CRE 6.1 115 0.989 0.014
MA 7.2 4.29 0.997 0.16
MMA 8.6 1.51 0.997 0.47
EMA 6.1 0.867 0.997 0.032
3-HGA 10 0.753 0.998 0.50
2-HGA 5.4 9.65 0.993 0.12
OA 5.6 0.297 0.998 0.026
MCA 5.7 0.305 0.998 0.029
LPC 20:0 8.2 0.0809 0.998 0.032
LPC 22:0 7.5 0.0159 0.998 0.00018
LPC 24:0 9.3 0.0294 0.991 0.0051
LPC 26:0 14 0.0264 0.992 0.010
Ile 11 9.85 0.994 0.050
aIle 11 0.347 0.992 0.065