表 2. Preliminary evaluation results for accuracy of monitoring instrument.
监测器准确性初步评估结果
| 志愿者 1 | 志愿者 2 | 志愿者 3 | ||||||||
|
G标准/
(mmol·L–1) |
G检测/
(mmol·L–1) |
相对误差
e (%) |
G标准/
(mmol·L–1) |
G检测/
(mmol·L–1) |
相对误差
e (%) |
G标准/
(mmol·L–1) |
G检测/
(mmol·L–1) |
相对误差
e (%) |
||
| 7.00 | 7.18 | 2.57 | 6.70 | 6.98 | 4.18 | 6.40 | 6.58 | 2.81 | ||
| 6.10 | 5.82 | 4.59 | 9.00 | 9.35 | 3.89 | 6.00 | 6.29 | 4.83 | ||
| 5.00 | 5.23 | 4.60 | 11.10 | 10.81 | 2.61 | 5.90 | 5.58 | 5.42 | ||
| 5.60 | 5.89 | 5.18 | 11.70 | 11.44 | 2.22 | 5.50 | 5.67 | 3.09 | ||
| 5.20 | 5.35 | 2.88 | 10.60 | 11.36 | 7.17 | 5.10 | 5.38 | 5.49 | ||