Skip to main content
. 2023 Feb 17;13(9):5900–5907. doi: 10.1039/d2ra07950j

Determination of glucose in blood and glucose oral liquid samples.

Samples (n = 3) This method (mM) Commercial glucose meter (mM)
Found Mean RSD Found Mean RSD
Blood sample 4.81 5.11 ±7.6% 4.94 4.95 ±3.6%
5.32 5.02
Glucose oral liquid 5.21 72.1 ±6.4% 4.90 71.6 ±4.7%
74.2 71.6
72.1 72.4
69.9 70.9