Table 1. Analytical Validation Parameters of nanoLC-HRMS Method for OXA and OXB Measurementa.

   
  intraday (N = 5)
interday (N = 15)
analyte matrix theoretical concentration (pg/mL) RE (%) CV (%) RE (%) CV (%)
OXA surrogate LLOQ 2.5 –3.3 7.8 1.9 12.8
LQC 7.5 –4.9 4.8 –1.5 6.6
MQC 120 0.4 2.6 0.3 2.1
HQC 200 0.6 1.8 –0.5 1.9
real LQC 57.7 –9.3 7.7 –8.8 14.0
HQC 157.7 –0.8 5.0 0.3 6.4
OXB surrogate LLOQ 2.5 0.0 10.3 2.0 9.6
LQC 7.5 0.1 4.6 –4.3 8.2
MQC 120 –7.6 2.4 –3.3 2.7
HQC 200 –2.1 3.3 –3.9 2.9
real LQC 28.2 –3.5 8.6 –5.6 9.7
HQC 128.2 1.6 2.6 –4.0 5.4
a

Intraday and interday assay precision (% CV) and accuracy (% RE) were evaluated in the surrogate matrix at 2.5 pg/mL (LLOQ), 7.5 pg/mL (lower quality control [LQC]), 120 pg/mL (middle quality control [MQC]), and 200 pg/mL (higher quality control [HQC]), and in real matrix spiked with orexins at 25 pg/mL (LQC) and 125 pg/mL (HQC).