. 2023 Feb 17;13:2868. doi: 10.1038/s41598-023-29973-3

Table 2.

Comparison of Elecsys ECLIA N-antigen measurements in EDTA/Heparin plasma without vs with addition of the extraction buffer. (95% confidence intervals).

Sample type EDTA without vs with extraction buffer Heparin without vs with extraction buffer
Number of samples 10 10
Slope 1.03 (0.96 to 1.1) 0.99 (0.86 to 1.1)
Y-intercept − 0.046 (− 0.31 to 0.22) 0.14 (− 0.41 to 0.68)
Spearman r  > 0.99  > 0.99