Postoperative audiological assessment (functional gain).


Sound-field PTA
SRT
WRS
Audio processor With BAHA (dB HL) mean ± SD Without BAHA (dB HL) mean ± SD Gain with BAHA (dB HL) mean With BAHA (dB HL) median±IQ Without BAHA (dB HL) median±IQR Gain with BAHA (dB HL) median With BAHA (%) median ± IQR Without BAHA (%) median±IQR
Baha® 5 (n = 14) 27.1 ± 6.3 55.5 ± 17.2 28.4 30.0 ± 10.0 52.5 ± 30.0 27.5 97.5 ± 5.0 87.5 ± 15.0
Baha® 5 Power (n = 6) 32.3 ± 6.9 61.83 ± 10.4 29.5 40.0 ± 8.8 67.5 ± 27.5 27.5 95.0 ± 6.3 85.0 ± 21.3
Baha® 5 SuperPower (n = 2) 36.3 ± 1.8 68.1 ± 9.7 31.8 42.5 ∗ 70.0 ∗ 27.5 90.0 ∗ 77.5 ∗
Overall (n = 22) 29.4 ± 6.8 58.4 ± 15.2 29 ǂ 40.0 ± 11.3 62.5 ± 30.0 22.5 Ɨ 95.0 ± 6.25 Ɨ 85.0 ± 15.0 Ɨ

Legend: PTA – Pure tone average; SRT – Speech recognition threshold; WRS - word recognition score; HL – Hearing level; SD – Standard deviation; IQR – Interquartile range.

IQR could not be calculated (n = 2).

ǂ Paired t-test: p = 0.002.

Ɨ Wilcoxon's signed-rank test: p < 0.001.