Table 4.

CNR of the optical nerve and inferior rectus muscle.

Scanner CTDIvol (mGy) Reconstruction technology Kernel Optical nerve CNR (SD) Inferior rectus muscle CNR (SD)
Reference scanner 2.66 FBP standard 2.80 (1.72) 2.36 (1.14)
VEO standard 10.18 (2.47) b 8.25 (2.91) a
NR40 16.36 (3.03) b 10.60 (4.45) b
Test scanner 1 2.65 FBP H30f 4.49 (0.75) 2.89 (0.60)
1.92 FBP H30f 3.98 (0.91) 4.19 (1.53)
Test scanner 2 2.65 FBP H30f 4.17 (0.87) 4.67 (3.25)
SAFIRE 3 J30f 5.12 (1.41) 5.47 (3.58)
SAFIRE 5 J30f 5.90 (1.91) 6.01 (3.64)
1.73 FBP H30f 3.40 (0.81) 3.25 (1.33)
SAFIRE 3 J30f 4.32 (1.08) 3.92 (1.69)
SAFIRE 5 J30f 5.08 (1.41) 3.25 (2.03)

CNR, contrast-to-noise ratio; FBP, filtered back projection; LDP, low dose protocol; SAFIRE, Sinogram Affirmed Iterative Reconstruction; SD, standard deviation.

a

Statistically significant (p < 0.05) difference to the reference scanner protocol FBP standard.

b

Statistically significant (p < 0.001) difference to the reference scanner protocol FBP standard.