Skip to main content
. Author manuscript; available in PMC: 2023 Dec 1.
Published in final edited form as: Med Phys. 2022 Nov 7;49(12):7373–7383. doi: 10.1002/mp.15979

Figure 5:

Figure 5:

Through-time temperature uncertainty estimates in the phantom acquired with an 18 cm field of view. Average and standard deviation of the values within the green bounding box are also reported. Aliasing from the water coupling medium introduces additional temperature uncertainty relative to the stationary state. The IBCM nearly completely removes the effects of aliasing on temperature uncertainty in the phantom. The asterisk indicates statistically significant difference in average temperature uncertainty.