Causes of revision.

Aetiology of arthritis COFAS type Preoperative coronal plane deformity (tibio-talar angle) Postoperative coronal alignment Initial lucency? Time to revision, mths Reason for revision Action
Osteoarthritis 4 8° valgus 10° varus No 5 Poor bone stock, Intraoperative fracture, lost anterior cortical contact and continued to dorsiflex Revised to INBONE
Post-traumatic 4 5° varus Nil initially
Hazy tibial lucency < 2 mm at 6/12 months
13 Aseptic loosening tibia Two stage revision to INBONE
Osteoarthritis 2 15° varus 2° varus No 4 Deep infection Two stage revision, awaiting IMBONE
Post-traumatic 2 12° varus 4° varus No, tibial radiolucency at 9 months 18 Aseptic loosening tibia Two stage revision, awaiting INBONE
Osteoarthritis 1 Yes 19 Aseptic loosening tibia Two stage revision, awaiting INBONE

COFAS, Canadian Orthopaedic Foot and Ankle Society.