Skip to main content
. 2023 Feb 16;11(2):585. doi: 10.3390/biomedicines11020585

Figure 5.

Figure 5

Difference between set powers (Pset) and measured powers (Pm) of the three K-lasers (K1, (ae); K-2, (fj); K-3, (ko)) operated in pulsed wave (PW) mode at varying repetition rates. The lasers were measured using each wavelength individually: (a,f,k), 800 nm; (b,g,l), 905 nm; (c,h,m), 970 nm; they were also measured using all diodes simultaneously: (d,i,n), all diodes. In addition, they were measured in intense super pulse (ISP) mode, which is also using all diodes but in a different pulsing mode: (e,j,o), ISP. Dashed horizontal lines at zero in each diagram illustrate ideal curves where Pm equals Pset.