Skip to main content
Micromachines logoLink to Micromachines
. 2023 Feb 2;14(2):374. doi: 10.3390/mi14020374

Review of Piezoelectric Micromachined Ultrasonic Transducers for Rangefinders

Jiong Pan 1, Chenyu Bai 1, Qincheng Zheng 1, Huikai Xie 1,2,*
Editor: Wei Li
PMCID: PMC9961946  PMID: 36838074

Abstract

Piezoelectric micromachined ultrasonic transducer (pMUT) rangefinders have been rapidly developed in the last decade. With high output pressure to enable long-range detection and low power consumption (16 μW for over 1 m range detection has been reported), pMUT rangefinders have drawn extensive attention to mobile range-finding. pMUT rangefinders with different strategies to enhance range-finding performance have been developed, including the utilization of pMUT arrays, advanced device structures, and novel piezoelectric materials, and the improvements of range-finding methods. This work briefly introduces the working principle of pMUT rangefinders and then provides an extensive overview of recent advancements that improve the performance of pMUT rangefinders, including advanced pMUT devices and range-finding methods used in pMUT rangefinder systems. Finally, several derivative systems of pMUT rangefinders enabling pMUT rangefinders for broader applications are presented.

Keywords: pMUT, piezoelectric, rangefinder, range-finding, microelectromechanical system, MEMS

1. Introduction

Piezoelectric micromachined ultrasonic transducers, or pMUTs, are microelectromechanical system (MEMS) devices that convert electrical energy to mechanical energy or vice versa through the flexural vibration (or deflection) of a membrane composed of a thin piezoelectric film based on piezoelectric effects [1,2,3]. A system with a single or several pMUTs can be used to transmit [4], to receive [5], or to both transmit and receive ultrasound signals (e.g., a pMUT imager [6,7], flowmeter [8], or rangefinder [9,10]), as illustrated in Figure 1. This article is focused on pMUT rangefinders. A pMUT rangefinder takes advantage of the signal transmitting and receiving capability of pMUTs to obtain a range between the target and the device [11], containing a single pMUT, multiple pMUTs, or an array of pMUTs to perform both ultrasonic transmitting and receiving. A pMUT rangefinder is only responsible for finding the range of a target but not the angle and direction of the target; derivative systems made of groups of pMUT rangefinders are able to perform more complicated tasks such as target positioning and object detection based on a group of range data [12,13]. It should be noted that pMUT rangefinder systems are different from pMUT imaging systems in that the objectives of pMUT imaging systems are to acquire the spatial distribution or three-dimensional (3D) structural information of a target but not the distances between the pMUT and the target points [14].

Figure 1.

Figure 1

Different applications of pMUTs.

pMUT rangefinders are promising in the relatively short-range (<10 m) detection for mobile use. Similar to other ultrasonic rangefinders, pMUT rangefinders are competitive to optical (light-based) rangefinder systems in relatively short-range detection because the speed of sound is much smaller than that of light, and they do not require high-speed electronic devices and their power consumptions are typically much lower. For instance, the power consumption of an optical rangefinder is greater than 1 W [13], while that of a pMUT rangefinder system can be only a few μW. It has been reported that a pMUT rangefinder can detect over a 1 m range under 16 μW of power consumption [15]. In addition, optical rangefinder systems are easily interfered with by sunlight or surrounding artificial lights, which may heavily affect range accuracy, limiting their use [16].

Secondly, compared to other ultrasonic rangefinders, pMUT rangefinders have their advantages. Currently, most commercially available ultrasonic rangefinders are made of bulk piezoelectric ultrasonic transducers due to their high output power [17], but they suffer from a series of drawbacks, including poor acoustic impedance matching [18] and large size [19]. Although incorporating acoustic matching layers can improve impedance matching [20], the bandwidth is limited, and it may not be practical in specific cases [21,22]. In contrast, pMUTs can have miniaturized size [23,24] and better acoustic matching [22,25], so a higher transmitting efficiency can be achieved. Furthermore, the small size of pMUTs not only enables pMUT rangefinders to be integrated with electronics, which saves space and energy for mobile use but also enables the fabrication of large ultrasonic transducer arrays, helping to realize more complicated systems [26,27,28]. Compared to ultrasonic rangefinders based on capacitive micromachined ultrasonic transducers (cMUTs) [29,30], pMUT rangefinders have a larger ultrasound output, lower bias voltage, and larger detectable range [31,32]. Hence, the low power consumption, relatively high output pressure, and integrated property of pMUT rangefinders make them promising for implementation into mobile devices for range-finding.

pMUT rangefinders were developed after bulk piezoelectric ultrasonic rangefinders, and the last decade witnessed the rapid development of pMUT rangefinders. At first, a single pMUT device with an AlN-based circular membrane was used to form a pMUT rangefinder system in 2010 [33]. After that, pMUT rangefinders with pMUT arrays, different types of pMUT device structures, and piezoelectric materials were developed to enhance the performance of pMUT rangefinder systems [34]. In addition to the advancements of pMUT devices, different range-finding methods are proposed [35]. Furthermore, various derivative systems based on pMUT rangefinders have been demonstrated, such as target positioning and object detection systems [16]. pMUT rangefinders are promising but still relatively new.

The main purpose of this article was to provide a state-of-the-art review of rangefinders based on pMUTs and a future outlook as well. Section 2 explains the principle of pMUT rangefinders and analyzes the performance of pMUT rangefinders. Section 3 reviews how the performance of pMUT rangefinders has improved through the advancement of pMUT devices. Section 4 reviews advanced pMUT rangefinders based on improving the range-finding method. Section 5 briefly introduces derivative systems of pMUT rangefinders. Section 6 summarizes this work and provides future outlooks of pMUT rangefinders.

2. Principle of pMUT Rangefinders

The working principles of pMUT rangefinders and the main factors affecting their performance are analyzed in this section to provide a foundation for pMUT rangefinder mechanism analysis and performance evaluation. In Section 2.1, the working mechanism and the signal propagation process of pMUT rangefinders are described. In Section 2.2, the analysis of pMUT rangefinder performance is given, and a quantitative relationship between the range and range error that determines the performance of pMUT rangefinders is deduced. After that, the main factors affecting the performance of pMUT rangefinders are analyzed in Section 2.3.

2.1. The Working Mechanism of pMUT Rangefinders

The block diagram of a pMUT rangefinder system is shown in Figure 2, where an electric transmitting signal from a signal generator is converted to an acoustic signal by the transmit mode of a pMUT device (①). The acoustic signal is reflected by the target (②), received by the receive mode of the pMUT device (③), which is converted back to the electric form (④) and picked up by an oscilloscope (⑤). Finally, the target range is given by data processing (⑥).

Figure 2.

Figure 2

Block diagram of a pMUT rangefinder system. The signal shown in the oscilloscope is from [36] (Reproduced with permission from [36], Copyright 2018, MDPI AG).

For traditional pMUT rangefinder systems, a single circular pMUT device is used to transmit and receive signals with a working frequency f, where the waveforms of the transmitting and receiving signals are pulses and echoes, and the system is based on the time-of-flight (TOF) method (also named the pulse-echo method) to find the range of the target. The threshold TOF method is the most commonly used one for pMUT rangefinders due to its simple implementation, where the TOF is measured by the time interval between the starting point and the time at which the leading edge of the echo reaches the amplitude threshold [37], as shown in Figure 3. The range between the pMUT and the target is given by [38]

R=cTOF2 (1)

where c is the sound speed.

Figure 3.

Figure 3

Illustration of the threshold TOF method. (a) The transmitting signals. (b) The receiving signal. Reproduced with permission from [39], Copyright 2019, IEEE.

Let us look at the signal propagation processes. The acoustic surface pressure amplitude Ptx generated by the transmit mode of the pMUT is proportional to the driving voltage Vd to excite the transmission mode of the pMUT, i.e.,

Ptx=GtVd (2)

where Gt is the transmitting sensitivity of the pMUT. The signal is spread spherically, reflected by the target, and received by the pMUT. For far-field radiation, when acoustic absorption is neglected, the on-axis acoustic pressure function p(r,t) at a distance r from a circular-membrane pMUT and a time, t, is given by [4,40,41]

p(r,t)=jρaccu0R0rej(ωtkr) (3)

where j is the imaginary unit, ρac is the mass density of the medium, u0 is the average velocity of the membrane, R0 is the Rayleigh distance, ω=2πf is the angular frequency, and k is the wave number. In Equation (3), the amplitude of the term “ρaccu0” is equal to Ptx [41]. Equation (3) is usually used to calculate the sound pressure level [4] or the transmitting sensitivity [42] of a pMUT in its transmit mode since Ptx in Equation (2) is not directly measurable and must be derived using Equation (3). For the cases in which acoustic absorption is not neglectable (e.g., in-air range-finding) and the acoustic gain Gac, which is related to the target geometry [31], must be considered, Equation (3) is changed to:

p(r,t)=jρaccu0R0rGaceαrej(ωtkr) (4)

where α is the absorption coefficient, which is used to evaluate the energy losses along the acoustic propagation path in the medium. Note that α is dependent on the properties of the medium, including the viscosity, thermal conductivity, mass density, etc., and is positively related to the ultrasound frequency [43].

Due to the round-trip propagation between the pMUT and the target, the amplitude of the received pressure at the pMUT, Prx, is obtained by setting r=2R in Equation (4), where R is the target range, i.e.,

Prx=PtxR02RGace2αR (5)

The received acoustic signal is converted to the electric signal by the pMUT in the receive mode, and the received voltage is given by:

Vrx=GrPrx (6)

where Gr is the receiving sensitivity of the pMUT. The signal propagation processes are listed in Table 1 in order.

Table 1.

The signal form alternation during the signal propagation process.

Number Signal Transmission Process Corresponding Signals
Acoustic pressure generated by the pMUT transmit mode Ptx=GtVd
Acoustic pressure received Prx=PtxR02RGace2αR
Electric signal generated by the pMUT receive mode Vrx=GrPrx

The relationship between the transmitting and receiving signals is described by Equations (2), (5) and (6), which will be used to analyze the performance of a pMUT rangefinder.

2.2. The Performance Evaluation of pMUT Rangefinders

A pMUT rangefinder is used for range-finding, so the maximum measurement range, Rmax, is the main criterion to evaluate its performance. Meanwhile, with the increase in the target range, the received pressure Prx is reduced and Vrx is decreased as well, according to Equations (5) and (6), and the accuracy of the range detection and the reliability of the range-finding result is lowered. Therefore, both Rmax and the range-finding accuracy must be considered for the performance comparison of different pMUT rangefinders.

The range-finding accuracy of a pMUT rangefinder is determined by multiple aspects, including Vrx, noises and interferences, the bandwidth, and the range-finding method. A large receiving signal amplitude makes the feature points, especially the leading edge, clearer, so the TOF determination and range detection can be more accurate. Vrx is related to both the electrical and acoustic energy conversion processes according to Equations (2) and (6). With the increasing distance of the target, Vrx or Prx decreases due to two main effects: one is that the signal is spread spherically and only a very small fraction of the energy is reflected by the target and received by the pMUT, and the other is acoustic absorption in the medium, according to Equation (5). Noises and interferences in a pMUT rangefinder system blur the signal and thus reduce the range-finding accuracy. Primary noises and interferences include the thermal noise caused by the random motion of the medium molecules received by the pMUT, the thermal noise of the piezoelectric element, the noises from the receiver circuit, the crosstalk between adjacent pMUT elements, the environmental reflections, and the possible multipath fading if the signal is continuous [44,45,46]. The signal-to-noise ratio (SNR) describes and evaluates the effects of Vrx as well as the noise and interference level on the accuracy. Combining Equations (2), (5) and (6) with the definition of SNR produces [37]:

SNR=SN=12Vrx2n2¯=Vd22n2¯(GtGrGacR0)2e4αR4R2 (7)

where S is the signal power, N is the noise power, and n2¯ is the mean-square of the noise voltage. Apart from the factors described by SNR, the range-finding accuracy is also affected by the pMUT bandwidth (BW) and the range-finding method. Ideally, the receiving signal is not spread and/or distorted compared to the transmitting signal, and the feature points, such as the leading edge, are sharp. However, due to the narrowband frequency response of the pMUT’s membrane [11], the rise and fall times of the receiving signal depend on the BW, and the feature point positioning error is introduced by the reduction in the signal rise and fall slope. In addition, pMUT rangefinder systems using different range-finding methods have different signal waveforms or the methods to derive the target range from the signals are different, so the accuracies of the ranging results are not the same. To include those multiple aspects influencing the range-finding accuracy, a range error is defined and commonly used for range-finding accuracy evaluation [37]. For the threshold TOF method, which is the most commonly used for pMUT range-finding, the root-mean-square (RMS) range error, δR, is given by [47]

δR=c2BW1(2SNR)12 (8)

In this paper, the RMS range error is abbreviated as the range error. The accuracy of a pMUT rangefinder system is revealed by its maximum range error δRmax. Overall, in this paper, the maximum range Rmax and the maximum range error δRmax are mainly considered to evaluate the performance of pMUT rangefinders.

In Equation (8), the bandwidth for a single-mode membrane is given by BW=f0Q, where f0 is the resonant frequency and Q is the quality factor [48,49]. Combining Equations (7) and (8) produces:

 δR=Qcn2¯GtGrGacR0f0VdRe2αR (9)

According to Equations (7) and (9), when R increases, SNR almost exponentially decreases while δR almost exponentially increases. This prediction is in good agreement with the experimental results of pMUT rangefinders, as shown in Figure 4. If the range error is large, the range-finding result given by the pMUT rangefinder is inaccurate. Plugging δRmax and Rmax into Equation (9) yields

 δRmax=Qcn2¯GtGrGacR0f0VdRmaxe2αRmax (10)

Figure 4.

Figure 4

(a) δR versus R. Reproduced with permission from [33], Copyright 2010, IEEE. (b) SNR versus R. Reproduced with permission from [31], Copyright 2022, MDPI AG.

When δR is required to be less than the maximum allowed range error δRmax, the Rmax that a pMUT rangefinder can achieve is determined by multiple factors that will be discussed in Section 2.3.

2.3. Main Factors Affecting pMUT Rangefinder Performance

According to Equation (10) and the previous analysis, with the enhancements of Gt and Gac, an increased Rmax can be achieved with fixed δRmax, where Gac is related to the target size and the directivity that can be changed by different pMUT structures [50], PCB baffle sizes [51], etc. A commonly used way to increase Gac by making the signal more directional is to add a horn on the pMUT [52]. In addition, with the increase in Gr, the receiving electric signal amplitude is increased with the side effect that the noise caused by the acoustic thermal noise is also increased. However, since acoustic noise is only one of the noise sources, the SNR is still largely increased with the enhancement of Gr and thus Rmax can be increased accordingly. Gt, Gr, and Gac can be enhanced by the pMUT device improvement.

Admittedly, the Rmax can be increased or the δRmax can be decreased by increasing the value of the driving voltage Vd according to Equation (10). However, higher driving voltage means higher power consumption and even nonlinearity [13], so both high performance and a moderate driving voltage should be achieved. In addition, the Rmax is positively related to the Rayleigh distance R0 with a fixed δRmax in Equation (10), and R0 is given by [40]:

R0=Aλ=fAc (11)

where A is the membrane surface area, λ is the wavelength, and f is the frequency of the signal transmitted by the pMUT and is referred to as the working frequency in this paper. According to the above equation, increasing A leads to a higher Rayleigh distance R0, and thus a larger Rmax with fixed δRmax, but the size and power consumption of the pMUT are also increased, and the resonant frequency is changed as well [53]. Therefore, the strategy to enhance the performance by directly changing the membrane area is rarely adopted.

Some factors have dual effects on the performance that should be properly designed, including the quality factor Q. Firstly, the increase in Q causes a decrease in the bandwidth, so Rmax is decreased or δRmax is increased, according to Equation (10). Secondly, the transmitting and receiving sensitivities Gt and Gr are also relevant to Q. A high Q means low acoustic and structural damping of pMUTs [54]. Therefore, with the increase in Q, the energy conversion efficiency is increased, and Gt and Gr are enhanced. Because of the dual effects of Q on the parameters of pMUTs, Q should be designed to be a moderate value to balance the high bandwidth and high sensitivity requirements for lower range error and larger detection range.

The working frequency f and the resonant frequency f0 also have multiple effects on pMUT rangefinder performance. f is the actual frequency of the pMUT driving signal while f0 is a property of the pMUT and is determined by the pMUT material and structure. In most pMUT rangefinder systems, to ensure the maximum vibration of the pMUT membrane, f should be set as approximately equal to f0 [15]. The signal amplitude reduction due to the difference between f and f0 causes the increase in δRmax or decrease in Rmax. According to Equation (10), for media with high absorption, the primary effect of f is acoustic absorption. An example of the relationship between the transmission loss and the range in a spread-only situation without absorption and in acoustic absorption situations with different frequencies is shown in Figure 5. With the decrease in f, the absorption coefficient α is decreased [40], so the transmission loss is reduced, and the Rmax is increased with fixed δRmax. In addition, with the reduction in f, the electric and mechano-acoustic impedances are decreased, so the receiving sensitivity Gr can be enhanced [54]. However, f should not be too low because the Rayleigh distance R0 is decreased with f reduction according to Equation (11), resulting in a higher range of error. Furthermore, since f0 should be close to f, with small f, the pMUT device has to be redesigned to have lower f0, which in turn causes lower bandwidth and higher range error. In addition, the quality factor Q may be also altered with the change in f0 [55]. Therefore, multiple effects of f and f0 must be considered during the design of the pMUT in order to obtain the optimal frequency of the pMUT rangefinder system in different scenarios. For the in-air range-finding, the working frequency should be moderately low due to the large absorption of air. For instance, a pMUT rangefinder first achieving more than a five-meter detection range with over 11.5 dB SNR was proposed by Yang et al. [31], where the working frequency was 66 kHz, which is lower than most previous designs. For in-liquid range-finding, the acoustic absorption is low, and the working frequency is usually in the MHz range [46] for higher bandwidth and larger Rayleigh distance.

Figure 5.

Figure 5

The transmission loss in a spread-only situation without absorption and in-air absorption situations with different frequencies at 20 °C and 50% air humidity. Reproduced with permission from [31], Copyright 2022, MDPI AG.

The mean-square of the noise voltage n2¯ and the absorption coefficient α negatively affect pMUT rangefinder performance. vn2¯ is determined by the noise and interference sources from the pMUT rangefinder and environment that are discussed in Section 2.2. α is determined by the working frequency f and properties of the medium that are discussed in Section 2.1. The reduction in n2¯ and/or α gives rise to higher Rmax or lower δRmax, according to Equation (10).

Overall, Table 2 lists the main factors affecting pMUT rangefinder performance and the trends of the maximum range Rmax with the increase in those factors when the maximum range error δRmax is fixed. Some of the factors could be improved by employing advanced pMUT devices and range-finding methods. Strategies to improve pMUT devices and range-finding methods for high-performance pMUT rangefinders are analyzed in detail in Section 3 and Section 4, respectively.

Table 2.

Main factors that affect the performance of pMUT rangefinders.

Name Symbol Main Effect *
The driving voltage Vd Rmax 
The working frequency f Dual effects **
The resonant frequency f0 BW  Rmax  **
The transmitting sensitivity Gt Rmax 
The receiving sensitivity Gr Rmax 
The acoustic gain Gac Rmax 
The quality factor Q Dual effects
The surface area of the membrane A Rmax 
The mean-square of the noise voltage n2¯ Rmax 
Absorption coefficient α Rmax 

* “↑” means that the increase in this factor will enhance the parameter, and “↓” means that the increase in this factor will lower the parameter. ** f should be approximately equal to f0 to achieve the maximum vibration, so their effects are not independent.

3. Advancements of pMUT Rangefinders Based on Improvements of pMUT Devices

A typical pMUT rangefinder system is made of a single pMUT device with a circular membrane and uses AlN as the piezoelectric material. This type of pMUT rangefinder has limited performance. In order to achieve the higher performance pMUT rangefinder systems, pMUT arrays, advanced pMUT device structures, and novel piezoelectric materials have been applied to pMUT rangefinders, which are discussed in detail in this section.

3.1. pMUT Rangefinders with pMUT Arrays

Using pMUT arrays to substitute a single pMUT for range-finding can help to achieve a higher maximum range or lower range error. As more pMUT elements are used for signal transmitting and receiving, the total transmitting and receiving acoustic signal amplitude are manyfold increased, so the SNR is increased, and the range error is largely reduced [13]. In addition, pMUT arrays with more element numbers can help to improve the directivity [56,57], so more acoustic energy is transmitted along the direction of the target and thus the receiving signal is enhanced, which in turn improves the maximum range and reduces the range error. In addition, since less energy along other directions is spread, the environmental reflection that affects the noise level is reduced, and thus n2¯ is lowered, as discussed in Section 2.2. For a one-dimensional (1D) array, assuming that the directivity with a single pMUT element is one, the directivity function D(θ) with an array of pMUT elements is given by [58]

D(θ)=sin(πNdλsinθ)Nsin(πdλsinθ) (12)

where N is the number of elements, d is the spacing between adjacent elements, and θ is the azimuth. The directivity function is represented by the normalized sound pressure and is shown in Figure 6. According to Figure 6, with the increase in the element number, the width of the main lobe is decreased, which means that the directivity is improved. The theory is also valid for two-dimensional (2D) arrays [59].

Figure 6.

Figure 6

The influence of the element number on the directivity. Reproduced with permission from [58], Copyright 2022, MDPI AG.

The types of pMUT array structures include line arrays (1D arrays) [60,61], polygon (e.g., hexagon) arrays [62,63], and square arrays [55,64], as shown in Figure 7. In many rangefinder systems using pMUT arrays, in order to keep the transmitter electrodes and the receiver electrodes in the same vibrational mode, respectively, all the transmitter electrodes are connected by metal bridges, and all the receiver electrodes are connected as well [65,66].

Figure 7.

Figure 7

(a) A 1D pMUT array. Reproduced with permission from [61], Copyright 2016, IEEE. (b) A hexagonal pMUT array. Reproduced with permission from [62], Copyright 2021, IEEE. (c) A 2D square array. Reproduced with permission from [23], Copyright 2013, MDPI AG. (d) A pMUT rangefinder system using a pMUT array.

Advanced pMUT array designs can improve the factors affecting pMUT performance, in which the density of pMUT elements and the mechanical or acoustic crosstalk between pMUT elements must be considered. By increasing the density of pMUT elements, the transmitting and receiving signal amplitude per unit area is enhanced. However, when reducing the spacing between pMUT elements to increase the element density, the crosstalk or cross-coupling between different elements must be carefully considered [23]. Strategies for minimizing the pMUT array crosstalk have been developed. For example, Yang et al. analyzed the mutual-coupling effect with different spacings in a square array of pMUTs and figured out the optimal spacing under a certain working frequency [23]. Moreover, Xu et al. proposed a resonant cavity design for a pMUT array to reduce crosstalk [67]. Vysotskyi et al. overviewed and compared acoustic isolation techniques for pMUT arrays concerning crosstalk reduction [68].

Table 3 lists the parameters of pMUT rangefinders with a single pMUT or different pMUT arrays. According to Table 3, pMUT arrays with large numbers of elements generally have high maximum ranges or high accuracies, which accords with the analysis. Note that the examples listed in Table 3 have different structures, piezoelectric materials, driving voltages, and δRmax or SNRmin. Thus, more control experiments of range-finding using single pMUTs and pMUT arrays with different element numbers need to be carried out in the future to further evaluate the contribution of pMUT arrays on the performance enhancement of pMUT rangefinder systems.

Table 3.

Summary of a single pMUT rangefinder or rangefinder arrays with different element numbers.

Element Number Rmax (m) ** δRmax (mm) ** SNRmin Vd (V) f (kHz) f0 (kHz) Ref.
1 element 0.45 1.3 6.5 214 214 [69]
7 receiver elements * 0.75 3.5 28 dB (at 0.5 m) 15 190 [52]
7 elements 0.5 ~0.67 *** 5 77.34 [62]
4 elements 2.4 12 dB 5 ~48 [70]
14 elements 6.8 11.5 dB 5 66 66 [31]

* [52] used a line of seven elements in an array as the receiver units. ** All the ranges and the corresponding range errors that were measured by the distance between the transmitter and the receiver were then divided by two to be converted to the reflection mode results for performance comparison. *** The range error reported in [62] is seen as the 3σ error, which is divided by three to be converted to the RMS errors and then divided by two to be converted to the reflection mode results in this paper.

3.2. pMUT Rangefinders with Advanced pMUT Device Structures

In addition to using pMUT arrays to enhance range-finding performance, pMUT rangefinders with advanced device structures have been exploited as well. For instance, some key parameters, including the transmitting and receiving sensitivities, are improved by optimizing the membrane shape of pMUT devices. Most conventional pMUT rangefinders use circular membrane pMUTs; nonetheless, some pMUT rangefinder designs have adopted square membranes for pMUTs recently, as shown in Figure 8. The vibrational modes of a pMUT are determined by the shape of the piezoelectric membrane, as shown in Figure 9. For the same device size, square-membrane pMUT units have a higher coverage area than circular ones [42], so the membrane vibration displacement of square-membrane pMUTs is largely higher than circular ones with the same driving voltage, which is proved by a comparison experiment in [71], as shown in Figure 10. Therefore, square membrane pMUT rangefinders have the potential for higher sensitivities Gt and Gr than current circular ones and achieve a larger detection range and lower range error. Chiu et al. reported a square-membrane pMUT rangefinder system [72] that achieved a higher maximum range (0.5 m) and lower range error (0.63 mm) than a comparable circular-membrane pMUT rangefinder (0.45 m maximum range with 1.3 mm error) [33]. In the future, the square-membrane pMUT structure may be combined with other strategies to further enhance pMUT rangefinder performance.

Figure 8.

Figure 8

(a) Comparison between the structure of a square-membrane pMUT (left) and a circular-membrane pMUT (right). Reproduced with permission from [71], Copyright 2018, IEEE. (b) A 3D model of square- and circular-membrane pMUTs. Reproduced with permission from [71], Copyright 2018, IEEE. (c) A pMUT rangefinder system using a square-membrane pMUT.

Figure 9.

Figure 9

(a) The shape of deflection of a circular membrane. Reproduced with permission from [39], Copyright 2019, IEEE. (b) The shape of deflection of a square membrane. Reproduced with permission from [36], Copyright 2018, MDPI AG.

Figure 10.

Figure 10

Tested vibration displacements of square- and circular-membrane pMUTs. Reproduced with permission from [71], Copyright 2018, IEEE.

Apart from the membrane shape change, Zhou et al. proposed a dual-electrode pMUT rangefinder system [15], as shown in Figure 11a. The pMUT transmitter has both inner and outer electrodes, and the driving electric signals of the electrodes have a 180°-phase difference. The electrodes with anti-phase driving signals control the displacement field of the membrane and improve the transmitting sensitivity. Therefore, the output acoustic pressure of the dual-electrode mode transmitter with anti-phase driving signals is considerably higher than the single electrode one, as shown in Figure 11b. Range-finding experiments were carried out, and the rangefinder system in the dual-electrode mode exhibited a 2.2 m maximum distance (i.e., a 1.1 m maximum range in the reflection target detection), which is 29.4% larger than the single electrode mode.

Figure 11.

Figure 11

(a) A pMUT rangefinder using a dual electrode transmitter driven by anti-phase electric signals. (b) The relationship between the acoustic pressure and the driving voltage in dual electrode and single electrode modes. Reproduced with permission from [15], Copyright 2017, Elsevier B.V.

There are other advanced pMUT device structures that are promising to be used for high-performance range-finding. For example, a bimorph structure is an effective way for pMUTs to enhance sensitivities, where the membrane of a pMUT contains two piezoelectric layers [73,74]. Because the number of piezoelectric layers is doubled, the vibration displacement and acoustic pressure of bimorph pMUTs are much larger than those of unimorph ones [75]. Shao et al. utilized a bimorph structure and developed a pMUT rangefinder derivative system to detect a target over 1 m away with a more than ±65° field of view [76], which was greater than the ±45° field of view of a unimorph design [13]. Moreover, other advanced structure designs, including the venting structure [50], dome shape [77], and unclosed membrane [78], may be further explored in order to extend the maximum range and lower the range error of pMUT rangefinders in the future.

Table 4 lists the parameters of typical pMUT rangefinders with different structures found in the literature. The maximum ranges and range errors of these pMUT rangefinders will also be illustrated in Figure 12.

Table 4.

Summary of pMUT rangefinders with different device structures.

Structure Features Rmax (m) * δRmax (mm) * SNRmin Vd (V) f (kHz) f0 (kHz) Ref.
Circular membrane 0.45 1.3 ~5.9 214 214 [33]
Circular membrane 0.65 ~0.28 ** ~27.5 dB 13 215 [11]
Square membrane 0.5 0.63 1.8 97 97, 96 [72]
Square membrane 1.4 ~0 dB 29 33 [36]
Dual electrode with anti-phase driving signals 1.1 12 dB 0.5 154 [15]

* All the ranges and the corresponding range errors that were measured by the distance between the transmitter and the receiver are divided by two to be converted to the reflection mode results for performance comparison. ** The 3σ error in [11] is divided by three to be converted to the RMS error and then divided by two to be converted to the reflection mode result.

3.3. pMUT Rangefinders with Novel Piezoelectric Materials

The transmitting sensitivity Gt and the receiving sensitivity Gr are mainly dependent on the piezoelectric materials of pMUT devices [79]:

{Gt|e31,f|Gr|e31,fε0ε33| (13)

where e31,f is the piezoelectric constant, ε0 is the vacuum permittivity, and ε33 is the relative dielectric constant in the z-direction. Therefore, the product of the sensitivities GtGr can be described by e31,f2ε0ε33, which is denoted as the figure of merit for a pMUT [80]. Since larger GtGr give rise to a smaller maximum range error δRmax or larger maximum range Rmax according to Equation (10), researchers have been working on choosing different piezoelectric materials and strategies to enhance GtGr by increasing the absolute value of e31,f or decreasing ε33 in order to enlarge the maximum range of pMUT rangefinders [81,82,83].

ZnO [84], PZT [85,86,87], and AlN [88,89,90] are widely used materials for pMUT devices with the advancement of thin-film deposition technologies. The drawback of ZnO is that it has a relatively high conductivity [91]. A PZT-based pMUT rangefinder has higher Gt due to its high piezoelectric coefficients, but its Gr is limited because of its high dielectric constants, so the GtGr or e31,f2ε0ε33 of PZT is not largely higher compared to that of AlN, as shown in Table 5 [92,93]. Furthermore, AlN is more compatible with CMOS fabrication than ZnO and PZT, which enables AlN pMUT rangefinders to be integrated in chips [92,94]. Overall, AlN is the most commonly used material for pMUT rangefinders.

However, AlN-based pMUT rangefinders have limited performance due to their low piezoelectric coefficients e31,f and high sensitivity to parasitic capacitance [91]. Since conventional PZT-based pMUTs cannot replace AlN-based pMUT rangefinders for their poor Gr, as it has been explained above, researchers applied various strategies to improve PZT films for advanced pMUT rangefinders with performance surpassing current AlN-based ones [95]. For example, S. Yoshida et al. developed a monocrystalline Pb(Mn1/3,Nb2/3)O3-PZT (PMnN-PZT) epitaxial thin film pMUT rangefinder [15]. According to Table 5, PMnN-PZT has a higher e31,f2ε0ε33, so it has a large GtGr. Therefore, the maximum range of PMnN-PZT pMUT rangefinders is increased to 2.2 m, i.e., 1.1 m in the reflection mode. Another example is the design of the single-crystal-PZT pMUT rangefinder developed by Luo et al. in 2020 [70], which has a maximum range of 2.4 m in the reflection mode. The e31,f2ε0ε33 of single-crystal PZT is not only obviously higher than conventional PZT but is also more than that of PMnN-PZT, according to Table 5. Furthermore, the single-crystal-PZT pMUT rangefinder in [70] has lower resonant frequencies (40~50 kHz) than the PMnN-PZT (>150 kHz) rangefinder in [15]. Since moderately low resonant frequencies are favorable for in-air pMUT rangefinders with long-range detection according to Section 2.3, the single-crystal-PZT-based pMUT rangefinder in [70] achieves higher Rmax.

Table 5.

Comparison of piezoelectric materials for pMUT rangefinders. Reproduced with permission from [70], Copyright 2021, IEEE.

Materials e31,f (C⋅m2) ε33 e31,f2ε0ε33 (Gpa)
AlN [96,97] −1 10 10.8
PZT (2 μm) [27] −13.1 854 22.7
PMnN-PZT [15] −14 ~250 88.6
Single-crystal PZT [70] −16~−24 308 93.7~211
20% ScAlN [98,99] −1.6 12 24.1

Apart from totally changing the piezoelectric material from AlN to other ones, other research has focused on doping AlN with suitable materials such as Scandium (Sc) to achieve a high piezoelectric coefficient [100,101]. This strategy balances the objectives that keep the advantages of AlN and increase the absolute value of e31,f. Yang et al. demonstrated a pMUT rangefinder array based on AlScN film, and a long-range detection with a maximum range of 6.8 m and SNR 11.5 dB was achieved [31].

Table 6 lists the parameters of pMUT rangefinders with different piezoelectric materials. According to Table 6, pMUT rangefinders based on PMnN-PZT, single-crystal PZT, and AlScN were observed to have obviously higher detection ranges than those on AlN and PZT. In particular, the AlScN pMUT rangefinder reaches the largest detection range (6.8 m), and the PMnN-PZT pMUT rangefinder in [15] realizes a long-range measurement of 1.1 m with very low power consumption (0.5 V driving voltage and about 16 μW), while a counterpart AlN pMUT rangefinder needs 10 V to reach the comparable Rmax.

Table 6.

Summary of pMUT rangefinders with different piezoelectric materials.

Piezoelectric Material Rmax (m) * δRmax (mm) * SNRmin Vd (V) f (kHz) f0 (kHz) Ref.
AlN 0.45 1.3 6.5 214 214 [69]
PZT 0.0175 0.1225 ** 1.5 ~30 ~30 [102]
PMnN-PZT 1.1 12 dB 0.5 154 [15]
Single-crystal PZT 2.4 12 dB 5 ~48 [70]
AlScN 6.8 11.5 dB 5 66 66 [31]

* All the ranges and the corresponding range errors that are measured by the distance between the transmitter and the receiver are divided by two to be converted to the reflection mode results for performance comparison. ** The error was less than 0.7% in [102], so δRmax of [102] was estimated by 0.7%Rmax.

The performances of TOF pMUT rangefinders with different pMUT devices are collectively presented in Figure 12. Note that for the performance comparison between two individual rangefinders, both Rmax and δRmax must be considered at the same time. For instance, the one with the same δRmax but higher Rmax, or the one with the same Rmax but lower δrmax, or the one with a higher Rmax and lower δRmax has a higher performance. In the figure, the colors of the markers stand for the working frequency f (f is substituted by the resonant frequency f0 for those pMUT rangefinders that did not explicitly provide f, and f is averaged for those with multiple working frequencies), ○ is a circular membrane pMUT, and ◊ is a square membrane pMUT. The hollow markers 🞉 and Inline graphic refer to those with no pMUT arrays used, while the solid markers ● and ⧫ refer to those with pMUT arrays as transmitters and/or receivers. Those pMUT rangefinders that used piezoelectric materials other than AlN are labeled. It can be seen in Figure 12 that moderately low working frequencies are desirable for pMUT rangefinders to achieve long-range detection, and the suitable working frequency range is approximately 40~200 kHz. Employing piezoelectric materials with high piezoelectric constants in pMUTs is the most effective means of extending the maximum range. In addition, pMUT arrays also contribute to the range error reduction or the maximum range enlargement. Square-membrane pMUT rangefinders have larger maximum ranges than their circular membrane counterparts, although the number of reported square-membrane pMUT rangefinders is relatively small.

Figure 12.

Figure 12

Performance comparison illustration of different TOF pMUT rangefinders. All the ranges and the corresponding range errors that are measured by the distance between the transmitter and the receiver are divided by two to be converted to the reflection mode results for performance comparison. 3σ errors are divided by three to be converted to the RMS errors. For those pMUT rangefinders that did not report their maximum range errors, δRmax is estimated by Equation (8), where the sound speed is set to be c= 343 m/s according to [103]. Data are from Ref. 1 ([33]), Ref. 2 ([11]), Ref. 3 ([52]), Ref. 4 ([13]), Ref. 5 ([55]), Ref. 6 ([15]), Ref. 7 ([36]), Ref. 8 ([102]), Ref. 9 ([39]), Ref. 10 ([62]), Ref. 11 ([72]), Ref. 12 ([70]), and Ref. 13 ([31]) in published year sequence.

4. Advancements of pMUT Rangefinders Based on Improvements of Range-Finding Methods

Section 3 describes the ways to enhance the performance of pMUT rangefinder systems by improving pMUT devices. Another form of performance enhancement is the improvement of the range-finding method. pMUT range-finding systems based on advanced TOF range-finding methods achieve high accuracy and wide detection range in long-range detection and are analyzed in Section 4.1. pMUT range-finding systems based on continuous wave (CW) range-finding methods achieve high accuracy in short-range detection and are analyzed in Section 4.2.

4.1. Advanced Time-of-Flight Range-Finding Methods for pMUT Rangefinders

4.1.1. TOF with Cross Correlation for Enhanced Accuracy

According to Section 2.1, traditional TOF pMUT rangefinder systems only use a single point to calculate the range, which may have a large range error because the attenuated received signals combined with considerable noises and interferences obscure the position of the leading edge. To address this problem, a cross-correlation TOF system was proposed, where signals were sampled, cross-correlation was performed, and the whole envelopes of the transmitting and receiving signals were compared and calibrated to determine the time of flight. Sampling sequences of the transmitting and receiving signals were cross-correlated [104], and the lag of the maximum cross-correlation peak, τmax, was given. Then, the distance between the transmitter and receiver, L, was given by [19,105]

L=(τmaxfsTOE)cLcal (14)

where fs is the sampling frequency, TOE is the time of the emission of the signal, and Lcal is a calibration constant. The range R between the pMUT rangefinder and the target can be obtained by combining Equation (14) with R=L/2.

In the cross-correlation method, the range result is determined not only by one point but by multiple feature points on the envelope, so the random noise can be mitigated and n2¯, in Equation (10), can be decreased, resulting in a lower range error and higher accuracy. Furthermore, in order to diminish the envelope distortion during energy conversion processes, the envelope shape of the transmitting signal in this system can be well designed to have enough smooth rising and falling slopes, as well as containing enough feature points [33]. A practical design of the envelope shape and the diagram of a pMUT rangefinder system applying the cross-correlation TOF method is shown in Figure 13 [33].

Figure 13.

Figure 13

Block diagram of a pMUT rangefinder system using the cross correlation TOF method. Reproduced with permission from [33], Copyright 2010, IEEE.

R. Przybyla et al. developed the first pMUT rangefinder system based on the cross-correlation TOF method as early as 2010 [33]. Many recent works also use cross correlation to obtain the detection range [31,106].

4.1.2. TOF with Ring-Down Suppression for Lower Blind Area

The maximum range is often the main parameter to be enhanced in most research on pMUT rangefinders in order to make pMUT rangefinders more competitive and to be applied to commercial range-finding use. Nevertheless, some recent research demonstrated strategies to lower the minimum detectable range. TOF pMUT rangefinder systems with a single pMUT device have a relatively large minimum detectable range due to the ring-down effect of the vibrating membrane, as shown in Figure 14. The ring-down effect is the main cause to extend the blind area of a pMUT rangefinder because the signal received during the ring-down stage is overlapped or distorted by the ring-down vibration, where the range error is largely increased, and the range-finding may fail [107]. The minimum detectable range of a TOF pMUT rangefinder can be defined by the blind area duration.

Figure 14.

Figure 14

Illustration of the ring down of a pMUT rangefinder. Reproduced with permission from [107], Copyright 2021, MDPI AG.

A simple way to avoid the blind area is to use two pMUT devices that are physically separated, where one of the pMUTs always works at the TX mode and the other always at the RX mode. However, the device size, readout circuit, and power consumption are largely increased if doubling the number of pMUT devices. Another strategy to lower the minimum detectable range in relatively short-range detection for pMUT rangefinders working in switching TX and RX modes is to redesign the transmitting signal to suppress the ring-down effect and reduce the blind area. For instance, Pala et al. proposed a method of ring-down suppression [108], in which the transmitting signal consisted of a number (Np) of normal cycles, such as the excitation signal and a number (Nn) of cycles that are 180° phase-shifted as the suppression signal, as shown in Figure 15a. The normal cycle and the 180° phase-shifted cycle signals counteract the vibration of the pMUT membrane and accelerate the vibration attenuation so that the ring-down time is reduced. The number of cycles Np and Nn must be properly determined. When Nn is not large enough, the counteracting effect is not adequate, while when Nn is too high, the additional 180° phase-shifted signal over-acts and excites the vibration of the pMUT membrane in its phase and frequency. In the experiment of [108], the optimal choice is Np=10 and Nn=6, as shown in Figure 15b, and the minimum detectable range (the blind area) is reduced from 13.91 cm to 12.16 cm. Furthermore, Wu et al. demonstrated a transfer function-based ring-down suppression method [107], which also achieved ring-down reduction through excitation signal suppression, but the suppression signal waveform was improved, as shown in Figure 15c, where the suppression signal was generated by a system including a transfer function and a simple proportion controller. The comparison experiment was carried out, and the minimum range (the blind area) was reduced by about 40% [107], as shown in Figure 15d. In the future, accuracy experiments of the ring-down suppression methods should be carried out to obtain the range error or SNR of the pMUT rangefinders using the methods in order to evaluate the accuracy of those methods in relatively short range detection compared with other designs.

Figure 15.

Figure 15

(a) Illustration of the 180° phase-shift ring-down suppression method [108]. Reproduced with permission from [108], Copyright 2021, IEEE. (b) The comparison experiment before and after suppression in [108]. Reproduced with permission from [108], Copyright 2021, IEEE. (c) The suppression signal of the ring-down suppression method is in [107]. Reproduced with permission from [107], Copyright 2021, MDPI AG. (d) The comparison experiment before and after suppression in [107]. Reproduced with permission from [107], Copyright 2021, MDPI AG.

4.2. Advanced Continuous Wave Range-Finding Methods for pMUT Rangefinders

4.2.1. MFCW pMUT Rangefinder System

The accuracy of TOF-based pMUT rangefinders is largely limited by the receiving signal amplitude which is heavily reduced by acoustic absorption [109]. Different from TOF methods that obtain the target range by amplitude information, continuous wave (CW) methods obtain the range by phase difference information, so they can achieve a higher accuracy than TOF methods [35]. Unlike TOF methods that the signals are pulses and echoes, CW methods transmit and receive continuous wave signals to find the range. The vibration modes of the signals in CW methods are steady, so they are less affected by bandwidth limitations. The change in the range-finding method from TOF methods to CW methods changes the relationship between δR and SNR [47], so the δR can be reduced even if Vd and n2¯ are not changed.

The simplest CW-based pMUT rangefinder system is to only use a sinusoidal wave as the signal. The phase detector can detect the phase difference between the transmitting and receiving signals, Δφ. The range is half of the distance L between the transmitter and the receiver and is given by [110]:

R=L2=cΔφ4πf (15)

where f is the frequency of the signal. However, since Δφ2π, Rc2f0, so Rmax=c2f0, which is very small. If the range is more than c2f0, phase ambiguity will occur. The simplest CW-based system cannot be used for pMUT rangefinders due to their low detection range. To address the problem of the detection range limitation, two frequency continuous wave (TFCW) systems have been proposed, where the signal contains two sinusoidal waves with slightly different frequencies f1 and f2 (|f1f2|=Δf), so the range is given by [111]:

R=L2=cΔφ4πΔf (16)

Since Δφ2π, Rmax=c2Δf, so Rmax is increased. The range error of a TFCW system is given by [47]:

δR=c2πBW1(2SNR)12 (17)

By comparing Equations (8) and (17), the range error of a TFCW system is π times lower than a threshold TOF system with the same SNR, bandwidth, and sound speed. However, there is a trade-off between Rmax and δRmax for the TFCW system. Although δR is decreased with the increase in Δf, Rmax is shortened when Δf is increased. To balance the parameters, Rmax cannot be too high and δRmax cannot be too low.

To solve this problem, multi-frequency continuous wave (MFCW) pMUT rangefinder systems are proposed, where three sinusoidal waves with frequencies of f1, f2, and f3 are transmitted simultaneously (f1>f2>f3). Define that Δfi,j=fifj is the frequency difference between the ith and jth waves, φi is the phase shift of the ith wave when it arrives at the receiver, and Δφi,j=φiφj is the phase shift difference between the ith and jth waves. The working mechanism of the MFCW method is illustrated in Figure 16, and the range R is given by [111]:

R=L2=c2(Int[Δφ1,2Δf1,32πΔf1,2]1Δf1,3+Int[Δφ1,3f12πΔf1,3]1f1+φ12πf1) (18)
Figure 16.

Figure 16

Illustration of the working mechanism of the MFCW method.

As long as Δf1,3 is sufficiently larger than Δf1,2, Rmax can be nearly independently increased with the decrease in Δf1,2, and the range error can be independently decreased with the increase in Δf1,3 and f1, where c/Δf1,3 determines the first-order resolution and c/f1 determines the second-order resolution. Thus, Rmax and δRmax are separately controlled [35]. As phenomenologically illustrated in Figure 17, for a TFCW system, when the period of the envelope is extended from Figure 17(a1) to Figure 17(a2), Rmax increases at the cost that the average slope of the envelope is also reduced, meaning that the resolution and the range error are decreased. In contrast, for an MFCW system, with the increase in the envelope’s period from Figure 17(b1) to Figure 17(b2), Rmax extends, and the average slope of the envelope (the spines with different heights) nearly does not change, which means that δRmax is approximately not increased. Furthermore, since MFCW systems use phase, not time-of-flight, to calculate the range, they are less affected by the large amplitude attenuation of the signal, which further enhances the accuracy of MFCW pMUT rangefinder systems.

Figure 17.

Figure 17

The illustration of TFCW waveforms (a1) Before period extension and (a2) After period extension, and MFCW waveforms (b1) Before period extension and (b2) After period extension.

In 2019, Chen et al. developed an MFCW-based pMUT rangefinder system and achieved a less than 0.0711 mm 3σ error (i.e., δR< 0.024 mm) when R< 0.1 m (for most TOF-based pMUT rangefinder systems, δR are not less than 0.4 mm) [35]. Although the detection range of this MFCW-based system is relatively low due to the maximum range limitation, MFCW-based pMUT rangefinder systems are promising to be used for high-accuracy short-range detection. The block diagram of an MFCW pMUT rangefinder system is shown in Figure 18. Different from a TOF system, the waveform generator must output three continuous waves with different frequencies [35].

Figure 18.

Figure 18

Block diagram of a MFCW pMUT rangefinder system. Reproduced with permission from [35], Copyright 2019, IEEE.

Table 7 lists the typical rangefinders with continuous wave methods. According to Table 7, MFCW rangefinders have larger Rmax and lower δRmax than the TFCW one. The conventional transducer-based range finder has a higher Rmax than the pMUT-based one, which means that the performance of the pMUT rangefinder should further be improved in the future to be competitive for commercial use.

Table 7.

Summary of rangefinders with continuous wave methods.

Device Method Rmax/m * δRmax/mm * Wave Frequencies Ref.
Conventional ultrasonic transmitter and receiver TFCW 0.1 1.5 39.85 and 40.6 kHz [112]
Conventional ultrasonic transducers MFCW 0.75 0.052 40, 39.9, 38.0 kHz [111]
pMUTs MFCW 0.3 ~0.024 for R < 0.1 m,
~0.61 for 0.1–0.3 m **
497, 496.8, 487 kHz for R < 0.1 m
492, 491.8, 490 kHz for 0.1–0.3 m
[35]

* All the ranges and the corresponding range errors that were measured by the distance between the transmitter and the receiver were divided by two to be converted to the reflection mode results for performance comparison. ** The 3σ errors in [35] are divided by three to be converted to the RMS errors.

4.2.2. MFPW pMUT Rangefinder System

MFCW-based pMUT rangefinder systems cannot achieve ideally high performance due to multipath reflections that are caused by the standing waves formed between the pMUT device and the target [45]. To eliminate this side effect, Zamora et al. proposed a novel system configuration for pMUT rangefinders named a multi-frequency pulsed wave (MFPW) system in 2021 [46], where three continuous waves are replaced by three pulses, each of which contains a single sinusoidal wave with several cycles. Additionally, the number of cycles, as well as the interval between distinct pulses, are well designed to make different pulses distinguishable and have a time-of-flight that is more than the total length of the transmitted signal, as shown in Figure 19. φ1, φ2, and φ3 are measured from these three pulses, respectively, and then Δφ1,2 and Δφ1,3 can be determined. The following steps to derive the range are similar as in the MFCW system described above. An MFPW system has a similar calculation process as an MFCW system, but it avoids interferences caused by multipath fading because discontinuous signals cannot form standing waves. The pMUT rangefinder system in [46] has an extremely low error: lower than 6.2 μm within a 3.5 mm transmitter-receiver distance.

Figure 19.

Figure 19

Illustration of the MFPW method. Reproduced with permission from [46], Copyright 2021, MDPI AG.

MFCW- and MFPW-based pMUT rangefinder systems are suitable to measure a short range (<several decimeters) because the Rmax is limited by the maximum phase shift difference of 2π. However, MFCW- and MFPW-based pMUT rangefinder systems have shown their extremely low range error when the range is short. Thus, they are competitive for short-range detection with high accuracy.

Overall, there are various kinds of pMUT rangefinder systems that are used for different applications. For long-range detection (>1 m), pMUT rangefinder systems should have moderately low working frequency according to Section 2.3 and utilize TOF range-finding methods. For short-range detection (<0.1 m) with the low-range-error requirement, pMUT rangefinder systems should utilize MFCW or MFPW range-finding methods.

5. Derivative Systems of pMUT Rangefinders

pMUT rangefinders described in Section 2, Section 3 and Section 4 can only find the range of a target but not the angle and direction of the target. Nevertheless, a group of pMUT rangefinders can be combined as an array to accomplish more complicated tasks such as target positioning (Section 5.1) and object detection (Section 5.2). These systems are named derivative systems of pMUT rangefinders because they are not simply pMUT rangefinder systems but are extensions of pMUT rangefinder systems.

5.1. pMUT Rangefinder Derivative Systems for Target Positioning

A pMUT rangefinder target positioning system can determine the position of a single point in a space so that the range and the angles of the point in a spherical coordinate frame can be measured. A target positioning system consists of N×N pMUT elements that are independently wired, and the time-of-flight of each element, τij, is separately calculated. For a target with the azimuths of θ and φ, the differences in the time-of-flight between neighboring elements of the pMUT array are given by [52]

{Δτx=dcsinθsinφΔτy=dcsinθcosφ (19)

where d is the adjacent element spacing (RNd), Δτx is the difference in the time-of-flight between neighboring elements along the x-axis, and Δτy is that along the y-axis. d has to be designed to be smaller than λ/2 to avoid grating lobes [113]. The target range R can be directly obtained by the τij data through Equation (1), and the azimuths θ and φ can be obtained by the known Δτx and Δτy data through Equation (19). Finally, the position of the target is given in Figure 20.

Figure 20.

Figure 20

Illustration of a pMUT rangefinder target positioning system.

5.2. pMUT Rangefinder Derivative Systems for Object Detection

pMUT rangefinder arrays can be utilized for object detection as well, where a group of range data of the points of an object [76] or a hand [16] is measured by the pMUT rangefinder array before data processing is used to detect the presence of the object or recognize the object features. An example of a pMUT rangefinder object detection system for gesture recognition is shown in Figure 21a.

Figure 21.

Figure 21

(a) A pMUT rangefinder object detection system for gesture recognition. (b) Block diagram of the pMUT rangefinder gesture recognition system. Reproduced with permission from [13], Copyright 2015, IEEE.

Because of the narrowband property of pMUTs, all the elements in a pMUT rangefinder array for object detection have the same working frequency. A CMOS circuitry capable of demodulation, frequency tracking, pulse generation, beamforming, and target tracking should be integrated with a pMUT rangefinder array to construct the system [13], as shown in Figure 21b, where beamforming is the most crucial part. Without deliberate beamforming, the echo signals reflected by multiple points of a target or multiple targets mix and superpose together, so one cannot distinguish a range difference from an angular difference by analyzing the signal differences of distinct elements for a pMUT rangefinder array. Hence, beam steering must be added so that the pMUT array only detects the range of a point in a certain direction at a time and then scans the direction [114,115]. The illustration of steering based on beamforming is shown in Figure 22. When measuring the range of target 1, phase delays are set to the elements of the pMUT array to compensate for the difference in the time-of-flight in the direction of target 1 (θ1), and then the received signals are summed together. Consequently, the received signal from θ1 is intensified, but other signals (e.g., in the direction of target 2 (θ2)) are not, so the range obtained this time is the range of target 1.

Figure 22.

Figure 22

Illustration of steering during beamforming.

6. Summary and Future Outlooks

This article discusses piezoelectric micromachined ultrasonic transducers (pMUTs) for range-finding. A brief introduction of pMUT rangefinders has been given, including their basic structures, working principles, advantages, and applications. Various designs for high-performance pMUT rangefinders have been discussed. pMUT arrays as well as advanced device structures (e.g., bimorph structure) and piezoelectric materials (e.g., AlScN) help to achieve high performance. Additionally, advanced range-finding methods (e.g., the MFPW method) are able to enhance the range-finding accuracy as well. Moreover, derivative systems of pMUT rangefinders that extend the use of pMUT rangefinders are briefly introduced.

With the maturing of advanced pMUT devices, pMUT rangefinders are promising as replacements for bulk piezoelectric transducers for commercial mobile range-finding because of their small size, low cost, and low power consumption. The performance of pMUT rangefinder systems should be further improved to enable pMUT rangefinders to be competitive consumer devices. Specifically, two main aspects should be focused on in the future: (1) the narrowband property of pMUTs makes their bandwidth generally lower than their counterparts, especially cMUTs, so electric and acoustic signal conversion processes inevitably have heavy distortions, and pMUT rangefinders are not competent to high-resolution 3D range-finding detection, especially in the air. The structural design, materials, and fabrication techniques of pMUTs have to be improved to increase their bandwidth, which, unfortunately, has not been extensively researched now. (2) It is still challenging to enable pMUT rangefinders with small sizes to generate a high output pressure that is comparable to bulk rangefinders. Therefore, in the future, pMUT rangefinders should be continuously improved through a deeper understanding of their principles, innovative structures, and/or new materials to reach a comparable output pressure as bulk rangefinders. Furthermore, high-performance range-finding methods such as frequency modulation and binary modulation that have not been applied to pMUT rangefinders are prospective to be researched for pMUT rangefinder systems for a particular use. In conclusion, the future development of pMUT rangefinders is challenging but promising.

Author Contributions

Methodology, conceptualization, and editing, J.P. and H.X.; writing—original draft preparation and figures preparation, J.P. and C.B.; writing—review and editing, J.P., C.B., Q.Z. and H.X.; supervision and project administration, H.X.; funding acquisition, H.X. All authors have read and agreed to the published version of the manuscript.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

Funding Statement

This research was funded by the Foshan Science and Technology Innovation Project (2018IT100252).

Footnotes

Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

References

  • 1.Qiu Y., Gigliotti J.V., Wallace M., Griggio F., Demore C.E., Cochran S., Trolier-McKinstry S. Piezoelectric micromachined ultrasound transducer (PMUT) arrays for integrated sensing, actuation and imaging. Sensors. 2015;15:8020–8041. doi: 10.3390/s150408020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Nazemi H., Antony Balasingam J., Swaminathan S., Ambrose K., Nathani M.U., Ahmadi T., Yameema B.L., Emadi A. Mass Sensors Based on Capacitive and Piezoelectric Micromachined Ultrasonic Transducers-CMUT and PMUT. Sensors. 2020;20:2010. doi: 10.3390/s20072010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Jiang X., Tang H.Y., Lu Y., Ng E.J., Tsai J.M., Boser B.E., Horsley D.A. Ultrasonic Fingerprint Sensor With Transmit Beamforming Based on a PMUT Array Bonded to CMOS Circuitry. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2017;64:1401–1408. doi: 10.1109/TUFFC.2017.2703606. [DOI] [PubMed] [Google Scholar]
  • 4.Wang H., Ma Y., Zheng Q., Cao K., Lu Y., Xie H. Review of Recent Development of MEMS Speakers. Micromachines. 2021;12:1257. doi: 10.3390/mi12101257. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Proto A., Rufer L., Basrour S., Penhaker M. Modeling and Measurement of an Ultrasound Power Delivery System for Charging Implantable Devices Using an AlN-Based pMUT as Receiver. Micromachines. 2022;13:2127. doi: 10.3390/mi13122127. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Lu Y., Tang H., Wang Q., Fung S., Tsai J.M., Daneman M., Boser B.E., Horsley D.A. Waveguide piezoelectric micromachined ultrasonic transducer array for short-range pulse-echo imaging. Appl. Phys. Lett. 2015;106:193506. doi: 10.1063/1.4921346. [DOI] [Google Scholar]
  • 7.Lu Y., Tang H., Fung S., Wang Q., Tsai J.M., Daneman M., Boser B.E., Horsley D.A. Ultrasonic fingerprint sensor using a piezoelectric micromachined ultrasonic transducer array integrated with complementary metal oxide semiconductor electronics. Appl. Phys. Lett. 2015;106:263503. doi: 10.1063/1.4922915. [DOI] [Google Scholar]
  • 8.van Neer P.L.M.J., Robers T., Volker A.W.F. A pMUT based flowmeter: A feasibility study; Proceedings of the 2013 IEEE International Ultrasonics Symposium (IUS); Prague, Czech Republic. 21–25 July 2013; pp. 1319–1322. [Google Scholar]
  • 9.Horsley D.A., Przybyla R.J., Shelton S.E., Goericke F.T., Eovino B.E., Alex M., Logan J. Piezoelectric Micromachined Ultrasonic Transducers for Range-Finding Applications; Proceedings of the 2021 5th IEEE Electron Devices Technology & Manufacturing Conference (EDTM); Chengdu, China. 8–11 April 2021; pp. 1–3. [Google Scholar]
  • 10.Robichaud A., Cicek P.-V., Deslandes D., Nabki F. Frequency Tuning Technique of Piezoelectric Ultrasonic Transducers for Ranging Applications. J. Microelectromechanical Syst. 2018;27:570–579. doi: 10.1109/JMEMS.2018.2831638. [DOI] [Google Scholar]
  • 11.Przybyla R., Flynn A., Jain V., Shelton S., Guedes A., Izyumin I., Horsley D., Boser B. A micromechanical ultrasonic distance sensor with >1 meter range; Proceedings of the 2011 16th International Solid-State Sensors, Actuators and Microsystems Conference; Beijing, China. 5–9 June 2011; pp. 2070–2073. [Google Scholar]
  • 12.Gijsenbergh P., Halbach A., Jeong Y., Torri G.B., Billen M., Demi L., Huang C.-H., Cheyns D., Rottenberg X., Rochus V. Characterization of polymer-based piezoelectric micromachined ultrasound transducers for short-range gesture recognition applications. J. Micromech. Microeng. 2019;29:074001. doi: 10.1088/1361-6439/ab1f41. [DOI] [Google Scholar]
  • 13.Przybyla R.J., Tang H.-Y., Guedes A., Shelton S.E., Horsley D.A., Boser B.E. 3D Ultrasonic Rangefinder on a Chip. IEEE J. Solid-St. Circ. 2015;50:320–334. doi: 10.1109/JSSC.2014.2364975. [DOI] [Google Scholar]
  • 14.Lu Y. Ph.D. Thesis. Mechanical and Aerospace Engineering, University of California; Oakland, CA, USA: 2015. Piezoelectric Micromachined Ultrasonic Transducers for Fingerprint Sensing. [Google Scholar]
  • 15.Zhou Z., Yoshida S., Tanaka S. Epitaxial PMnN-PZT/Si MEMS ultrasonic rangefinder with 2 m range at 1 V drive. Sens. Actuators A Phys. 2017;266:352–360. doi: 10.1016/j.sna.2017.09.058. [DOI] [Google Scholar]
  • 16.Przybyla R.J., Tang H.-Y., Shelton S.E., Horsley D.A., Boser B.E. 12.1 3D ultrasonic gesture recognition; Proceedings of the 2014 IEEE International Solid-State Circuits Conference Digest of Technical Papers (ISSCC); San Francisco, CA, USA. 9–13 February 2014; pp. 210–211. [Google Scholar]
  • 17.Kanda T., Makino A., Ono T., Suzumori K., Morita T., Kurosawa M.K. A micro ultrasonic motor using a micro-machined cylindrical bulk PZT transducer. Sens. Actuators A Phys. 2006;127:131–138. doi: 10.1016/j.sna.2005.10.056. [DOI] [Google Scholar]
  • 18.Wang T., Kobayashi T., Lee C. Highly sensitive piezoelectric micromachined ultrasonic transducer operated in air. Micro Nano Lett. 2016;11:558–562. doi: 10.1049/mnl.2016.0207. [DOI] [Google Scholar]
  • 19.Qiu Z., Lu Y., Qiu Z. Review of Ultrasonic Ranging Methods and Their Current Challenges. Micromachines. 2022;13:520. doi: 10.3390/mi13040520. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Toda M. New type of matching layer for air-coupled ultrasonic transducers. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2002;49:972–979. doi: 10.1109/TUFFC.2002.1020167. [DOI] [PubMed] [Google Scholar]
  • 21.Akasheh F., Myers T., Fraser J.D., Bose S., Bandyopadhyay A. Development of piezoelectric micromachined ultrasonic transducers. Sens. Actuators A Phys. 2004;111:275–287. doi: 10.1016/j.sna.2003.11.022. [DOI] [Google Scholar]
  • 22.Wang T., Lee C. Zero-Bending Piezoelectric Micromachined Ultrasonic Transducer (pMUT) With Enhanced Transmitting Performance. J. Microelectromechanical Syst. 2015;24:2083–2091. doi: 10.1109/JMEMS.2015.2472958. [DOI] [Google Scholar]
  • 23.Yang Y., Tian H., Wang Y.F., Shu Y., Zhou C.J., Sun H., Zhang C.H., Chen H., Ren T.L. An ultra-high element density pMUT array with low crosstalk for 3-D medical imaging. Sensors. 2013;13:9624–9634. doi: 10.3390/s130809624. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Herrera B., Pop F., Cassella C., Rinaldi M. Miniaturized PMUT-Based Receiver for Underwater Acoustic Networking. J. Microelectromechanical Syst. 2020;29:832–838. doi: 10.1109/JMEMS.2020.3018070. [DOI] [Google Scholar]
  • 25.Pala S., Lin L. Piezoelectric Micromachined Ultrasonic Transducers (pMUT) with Free Boundary; Proceedings of the 2020 IEEE International Ultrasonics Symposium (IUS); Las Vegas, NV, USA. 7–11 September 2020; pp. 1–4. [Google Scholar]
  • 26.Dausch D.E., Gilchrist K.H., Carlson J.R., Castellucci J.B., Chou D.R., von Ramm O.T. Improved pulse-echo imaging performance for flexure-mode pMUT arrays; Proceedings of the 2010 IEEE International Ultrasonics Symposium; San Diego, CA, USA. 11–14 October 2010; pp. 451–454. [Google Scholar]
  • 27.Muralt P., Ledermann N., Baborowski J., Barzegar A., Gentil S., Belgacem B., Petitgrand S., Bosseboeuf A., Setter N. Piezoelectric micromachined ultrasonic transducers based on PZT thin films. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2005;52:2276–2288. doi: 10.1109/TUFFC.2005.1563270. [DOI] [PubMed] [Google Scholar]
  • 28.Liu W., He L., Wang X., Zhou J., Xu W., Smagin N., Toubal M., Yu H., Gu Y., Xu J., et al. 3D FEM Analysis of High-Frequency AlN-Based PMUT Arrays on Cavity SOI. Sensors. 2019;19:4450. doi: 10.3390/s19204450. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Zhao L., Annayev M., Oralkan O., Jia Y. An Ultrasonic Energy Harvesting IC Providing Adjustable Bias Voltage for Pre-Charged CMUT. IEEE Trans. Biomed. Circuits Syst. 2022;16:842–851. doi: 10.1109/TBCAS.2022.3178581. [DOI] [PubMed] [Google Scholar]
  • 30.Wygant I.O., Kupnik M., Windsor J.C., Wright W.M., Wochner M.S., Yaralioglu G.G., Hamilton M.F., Khuri-Yakub B.T. 50 kHz capacitive micromachined ultrasonic transducers for generation of highly directional sound with parametric arrays. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2009;56:193–203. doi: 10.1109/TUFFC.2009.1019. [DOI] [PubMed] [Google Scholar]
  • 31.Yang H., Ji M., Xiu X., Lv H., Gu A., Zhang S. AlScN Film Based Piezoelectric Micromechanical Ultrasonic Transducer for an Extended Long-Range Detection. Micromachines. 2022;13:1942. doi: 10.3390/mi13111942. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Birjis Y., Swaminathan S., Nazemi H., Raj G.C.A., Munirathinam P., Abu-Libdeh A., Emadi A. Piezoelectric Micromachined Ultrasonic Transducers (PMUTs): Performance Metrics, Advancements, and Applications. Sensors. 2022;22:9151. doi: 10.3390/s22239151. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Przybyla R., Izyumin I., Kline M., Boser B., Shelton S. An ultrasonic rangefinder based on an AlN piezoelectric micromachined ultrasound transducer; Proceedings of the 2010 IEEE Sensors; Waikoloa, HI, USA. 1–4 November 2010; pp. 2417–2421. [Google Scholar]
  • 34.Zhou Z., Yoshida S., Tanaka S. Monocrystalline PMNN-PZT thin film ultrasonic rangefinder with 2 meter range at 1 volt drive; Proceedings of the 2017 19th International Conference on Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS); Kaohsiung, Taiwan. 18–22 June 2017; pp. 167–170. [Google Scholar]
  • 35.Chen X., Xu J., Chen H., Ding H., Xie J. High-Accuracy Ultrasonic Rangefinders via pMUTs Arrays Using Multi-Frequency Continuous Waves. J. Microelectromechanical Syst. 2019;28:634–642. doi: 10.1109/JMEMS.2019.2912869. [DOI] [Google Scholar]
  • 36.Alasatri S., Rufer L., Lee J.E.-Y. AlN-on-Si Square Diaphragm Piezoelectric Micromachined Ultrasonic Transducer with Extended Range of Detection; Proceedings of the Eurosensors 2018; Graz, Austria. 9–12 September 2018. [Google Scholar]
  • 37.Skolnik M. Introduction to Radar Systems. 3rd ed. McGraw-Hill; New York, NY, USA: 2001. [Google Scholar]
  • 38.Parrilla M., Anaya J.J., Fritsch C. Digital signal processing techniques for high accuracy ultrasonic range measurements. IEEE Trans. Instrum. Meas. 1991;40:759–763. doi: 10.1109/19.85348. [DOI] [Google Scholar]
  • 39.Suresh A., Mak K.L., Benserhir J., Lee J.E.Y., Rufer L. Air-coupled Ultrasonic Rangefinder with Meter-long Detection Range Based on a Dual-electrode PMUT Fabricated Using a Multi-user MEMS Process; Proceedings of the 2019 IEEE Sensors; Montreal, QC, Canada. 27–30 October 2019; pp. 1–4. [Google Scholar]
  • 40.Blackstock D.T. Fundamentals of Physical Acoustics. John Wiley and Sons; Hoboken, NJ, USA: 2000. [Google Scholar]
  • 41.Ledesma E., Zamora I., Uranga A., Barniol N. Tent-Plate AlN PMUT With a Piston-Like Shape Under Liquid Operation. IEEE Sens. J. 2020;20:11128–11137. doi: 10.1109/JSEN.2020.2995643. [DOI] [Google Scholar]
  • 42.Ledesma E., Zamora I., Torres F., Uranga A., Tzanov V., Barniol N., Marigó E., Soundara-Pandian M. Squared PMUT with Enhanced Pressure Sensitivities; Proceedings of the Eurosensors 2018; Graz, Austria. 9–12 September 2018. [Google Scholar]
  • 43.Kinsler L.E., Frey A.R., Coppens A.B., Sanders J.V. Fundamentals of Acoustics. 4th ed. Wiley Publications; Hoboken, NJ, USA: 2000. [Google Scholar]
  • 44.Wang H., Ma Y., Yang H., Jiang H., Ding Y., Xie H. MEMS Ultrasound Transducers for Endoscopic Photoacoustic Imaging Applications. Micromachines. 2020;11:928. doi: 10.3390/mi11100928. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Kuratli C., Qiuting H. A CMOS ultrasound range-finder microsystem. IEEE J. Solid-St. Circ. 2000;35:2005–2017. doi: 10.1109/4.890317. [DOI] [Google Scholar]
  • 46.Zamora I., Ledesma E., Uranga A., Barniol N. High Accuracy Ultrasound Micro-Distance Measurements with PMUTs under Liquid Operation. Sensors. 2021;21:4524. doi: 10.3390/s21134524. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Webster D. A pulsed ultrasonic distance measurement system based upon phase digitizing. IEEE Trans. Instrum. Meas. 1994;43:578–582. doi: 10.1109/19.310171. [DOI] [Google Scholar]
  • 48.Green E.I. The story of Q. Am. Sci. 1955;43:584–594. [Google Scholar]
  • 49.Jeong J.J., Choi H. An impedance measurement system for piezoelectric array element transducers. Measurement. 2017;97:138–144. doi: 10.1016/j.measurement.2016.10.053. [DOI] [Google Scholar]
  • 50.Rozen O., Block S.T., Shelton S.E., Horsley D.A. Piezoelectric micromachined ultrasonic transducer with increased output pressure via concentric venting rings; Proceedings of the 2015 Transducers-2015 18th International Conference on Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS); Anchorage, AK, USA. 21–25 June 2015; pp. 670–673. [Google Scholar]
  • 51.Nayak B., Gupta H., Roy K., Ashok A., Shastri V., Pratap R. An Experimental Study of the Acoustic Field of a Single-Cell Piezoelectric Micromachined Ultrasound Transducer (PMUT); Proceedings of the 2020 5th IEEE International Conference on Emerging Electronics (ICEE); New Delhi, India. 26–28 November 2020; pp. 1–4. [Google Scholar]
  • 52.Przybyla R.J., Shelton S.E., Guedes A., Krigel R., Horsley D.A., Boser B.E. In-Air Ultrasonic Rangefinding and Angle Estimation Using an Array of Aln Micromachined Transducers; Proceedings of the 2012 Solid-State, Actuators, and Microsystems Workshop Technical Digest; Hilton Head, SC, USA. 3–7 June 2012; pp. 50–53. [Google Scholar]
  • 53.Wang M., Zhou Y. Design of piezoelectric micromachined ultrasonic transducers (pMUTs) for high pressure output. Microsyst. Technol. 2016;23:1761–1766. doi: 10.1007/s00542-016-2929-9. [DOI] [Google Scholar]
  • 54.Dangi A., Pratap R. System level modeling and design maps of PMUTs with residual stresses. Sens. Actuators A Phys. 2017;262:18–28. doi: 10.1016/j.sna.2017.05.006. [DOI] [Google Scholar]
  • 55.Rozen O., Block S.T., Mo X., Bland W., Hurst P., Tsai J.M., Daneman M., Amirtharajah R., Horsley D.A. Monolithic MEMS-CMOS ultrasonic rangefinder based on dual-electrode PMUTs; Proceedings of the 2016 IEEE 29th International Conference on Micro Electro Mechanical Systems (MEMS); Shanghai, China. 24–28 January 2016; pp. 115–118. [Google Scholar]
  • 56.Herrera B., Pop F., Cassella C., Rinaldi M. AlN PMUT-based Ultrasonic Power Transfer Links for Implantable Electronics; Proceedings of the 2019 20th International Conference on Solid-State Sensors, Actuators and Microsystems & Eurosensors XXXIII (TRANSDUCERS & EUROSENSORS XXXIII); Berlin, Germany. 23–27 June 2019; pp. 861–864. [Google Scholar]
  • 57.Chen M., Zhang Q., Zhao X., Wang F., Liu H., Li B., Zhang X., Luo H. Design and analysis of piezoelectric micromachined ultrasonic transducer using high coupling PMN-PT single crystal thin film for ultrasound imaging. Smart Mater. Struct. 2021;30:055006. doi: 10.1088/1361-665X/abee37. [DOI] [Google Scholar]
  • 58.Li B., Sun C., Xin S., Luo M., Hei C., Du G., Feng A. Development of a 16-Channel Broadband Piezoelectric Micro Ultrasonic Transducer Array Probe for Pipeline Butt-Welded Defect Detection. Sensors. 2022;22:7133. doi: 10.3390/s22197133. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Liao W., Liu W., Rogers J.E., Usmani F., Tang Y., Wang B., Jiang H., Xie H. Piezeoelectric micromachined ultrasound tranducer array for photoacoustic imaging; Proceedings of the 2013 Transducers & Eurosensors XXVII: The 17th International Conference on Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS & EUROSENSORS XXVII); Barcelona, Spain. 16–20 June 2013; pp. 1831–1834. [Google Scholar]
  • 60.Cai J., Liu K., Lou L., Zhang S., Gu Y.A., Wu T. Increasing Ranging Accuracy of Aluminum Nitride Pmuts by Circuit Coupling; Proceedings of the 2021 IEEE 34th International Conference on Micro Electro Mechanical Systems (MEMS); Gainesville, FL, USA. 25–29 January 2021; pp. 740–743. [Google Scholar]
  • 61.Wang T., Kobayashi T., Yang B., Wang H., Lee C. Highly sensitive piezoelectric micromachined ultrasonic transducer (pMUT) operated in air; Proceedings of the 2016 IEEE 11th Annual International Conference on Nano/Micro Engineered and Molecular Systems (NEMS); Sendai, Japan. 17–20 April 2016; pp. 294–299. [Google Scholar]
  • 62.Cai G., Zhou X., Yi Y., Zhou H., Li D., Zhang J., Huang H., Mu X. An Enhanced-Differential PMUT for Ultra-Long Distance Measurement in Air; Proceedings of the 2021 IEEE 34th International Conference on Micro Electro Mechanical Systems (MEMS); Gainesville, FL, USA. 25–29 January 2021; pp. 899–902. [Google Scholar]
  • 63.Kusano Y., Luo G.-L., Horsley D.A. Spurious Mode Free Hexagonal PMUT Array; Proceedings of the 2019 20th International Conference on Solid-State Sensors, Actuators and Microsystems & Eurosensors XXXIII (TRANSDUCERS & EUROSENSORS XXXIII); Berlin, Germany. 23–27 June 2019; pp. 238–241. [Google Scholar]
  • 64.Sun S., Wang J., Zhang M., Yuan Y., Ning Y., Ma D., Niu P., Gong Y., Yang X., Pang W. Eye-Tracking Monitoring Based on PMUT Arrays. J. Microelectromechanical Syst. 2022;31:45–53. doi: 10.1109/JMEMS.2021.3125377. [DOI] [Google Scholar]
  • 65.Jung J., Kim S., Lee W., Choi H. Fabrication of a two-dimensional piezoelectric micromachined ultrasonic transducer array using a top-crossover-to-bottom structure and metal bridge connections. J. Micromech. Microeng. 2013;23:125037. doi: 10.1088/0960-1317/23/12/125037. [DOI] [Google Scholar]
  • 66.Jung J., Lee W., Kang W., Hong H., Song H.Y., Oh I.-y., Park C.S., Choi H. A top-crossover-to-bottom addressed segmented annular array using piezoelectric micromachined ultrasonic transducers. J. Micromech. Microeng. 2015;25:115024. doi: 10.1088/0960-1317/25/11/115024. [DOI] [Google Scholar]
  • 67.Xu T., Zhao L., Jiang Z., Guo S., Li Z., Yang P., Sun L., Luo G., Zhang L. Array Design of Piezoelectric Micromachined Ultrasonic Transducers With Low-Crosstalk and High-Emission Performance. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2020;67:789–800. doi: 10.1109/TUFFC.2019.2956181. [DOI] [PubMed] [Google Scholar]
  • 68.Vysotskyi B., Haouari R., Weekers B., Keulemans G., Torri G.B., Rochus V. A comparative study of acoustic isolation configurations for pMUT array-based applications; Proceedings of the 2021 22nd International Conference on Thermal, Mechanical and Multi-Physics Simulation and Experiments in Microelectronics and Microsystems (EuroSimE); St. Julian, Malta. 19–21 April 2021; pp. 1–5. [Google Scholar]
  • 69.Przybyla R.J., Shelton S.E., Guedes A., Izyumin I.I., Kline M.H., Horsley D.A., Boser B.E. In-Air Rangefinding With an AlN Piezoelectric Micromachined Ultrasound Transducer. IEEE Sens. J. 2011;11:2690–2697. doi: 10.1109/JSEN.2011.2157490. [DOI] [Google Scholar]
  • 70.Luo G.-L., Kusano Y., Horsley D.A. Airborne Piezoelectric Micromachined Ultrasonic Transducers for Long-Range Detection. J. Microelectromechanical Syst. 2021;30:81–89. doi: 10.1109/JMEMS.2020.3037298. [DOI] [Google Scholar]
  • 71.Liu X., Chen X., Le X., Xu Z., Wu C., Xie J. A High-Performance Square pMUT for Range-finder; Proceedings of the 2018 IEEE 13th Annual International Conference on Nano/Micro Engineered and Molecular Systems (NEMS); Singapore. 22–26 April 2018; pp. 106–109. [Google Scholar]
  • 72.Chiu Y., Wang C., Gong D., Li N., Ma S., Jin Y. A Novel Ultrasonic TOF Ranging System Using AlN Based PMUTs. Micromachines. 2021;12:284. doi: 10.3390/mi12030284. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Akhbari S., Sammoura F., Eovino B., Yang C., Lin L. Bimorph Piezoelectric Micromachined Ultrasonic Transducers. J. Microelectromechanical Syst. 2016;25:326–336. doi: 10.1109/JMEMS.2016.2516510. [DOI] [Google Scholar]
  • 74.Akhbari S., Sammoura F., Yang C., Mahmoud M., Aqab N., Lin L. Bimorph pMUT with dual electrodes; Proceedings of the 2015 28th IEEE International Conference on Micro Electro Mechanical Systems (MEMS); Estoril, Portugal. 18–22 January 2015; pp. 928–931. [Google Scholar]
  • 75.Shao Z., Pala S., Peng Y., Lin L. Bimorph Pinned Piezoelectric Micromachined Ultrasonic Transducers for Space Imaging Applications. J. Microelectromechanical Syst. 2021;30:650–658. doi: 10.1109/JMEMS.2021.3090374. [DOI] [Google Scholar]
  • 76.Shao Z., Peng Y., Pala S., Liang Y., Lin L. 3D Ultrasonic Object Detections with >1 Meter Range; Proceedings of the 2021 IEEE 34th International Conference on Micro Electro Mechanical Systems (MEMS); Gainesville, FL, USA. 25–29 January 2021; pp. 386–389. [Google Scholar]
  • 77.Nistorica C., Latev D., Mendoza J., Imai D. Modeling and characterization of a 3D-MEMS piezoelectric ultrasound transducer; Proceedings of the 2016 IEEE International Ultrasonics Symposium (IUS); Tours, France. 18–21 September 2016; pp. 1–4. [Google Scholar]
  • 78.Stoppel F., Mannchen A., Niekiel F., Beer D., Giese T., Wagner B. New integrated full-range MEMS speaker for in-ear applications; Proceedings of the 2018 IEEE Micro Electro Mechanical Systems (MEMS); Belfast, UK. 21–25 January 2018; pp. 1068–1071. [Google Scholar]
  • 79.Jung J., Lee W., Kang W., Shin E., Ryu J., Choi H. Review of piezoelectric micromachined ultrasonic transducers and their applications. J. Micromech. Microeng. 2017;27:113001. doi: 10.1088/1361-6439/aa851b. [DOI] [Google Scholar]
  • 80.Yoshida S., Hanzawa H., Wasa K., Tanaka S. Fabrication and characterization of large figure-of-merit epitaxial PMnN-PZT/Si transducer for piezoelectric MEMS sensors. Sens. Actuators A Phys. 2016;239:201–208. doi: 10.1016/j.sna.2016.01.031. [DOI] [Google Scholar]
  • 81.Luo G.-L., Kusano Y., Roberto M.N., Horsley D.A. High-Pressure Output 40 kHz Air-Coupled Piezoelectric Micromachined Ultrasonic Transducers; Proceedings of the 2019 IEEE 32nd International Conference on Micro Electro Mechanical Systems (MEMS); Seoul, Republic of Korea. 27–31 January 2019; pp. 787–790. [Google Scholar]
  • 82.Cheng C., Peters T., Dangi A., Agrawal S., Chen H., Kothapalli S.R., Trolier-McKinstry S. Improving PMUT Receive Sensitivity via DC Bias and Piezoelectric Composition. Sensors. 2022;22:5614. doi: 10.3390/s22155614. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 83.Ngoc Thao P., Yoshida S., Tanaka S. Fabrication and Characterization of PZT Fibered-Epitaxial Thin Film on Si for Piezoelectric Micromachined Ultrasound Transducer. Micromachines. 2018;9:455. doi: 10.3390/mi9090455. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 84.Li J., Ren W., Fan G., Wang C. Design and Fabrication of Piezoelectric Micromachined Ultrasound Transducer (pMUT) with Partially-Etched ZnO Film. Sensors. 2017;17:1381. doi: 10.3390/s17061381. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 85.Sadeghpour S., Puers R. Optimization in the Design and Fabrication of a PZT Piezoelectric Micromachined Ultrasound Transducer (PMUT); Proceedings of the Eurosensors 2018; Graz, Austria. 9–12 September 2018. [Google Scholar]
  • 86.Yu Q., Fan G., Ren W., Fan Q., Ti J., Li J., Wang C. PZT-Film-Based Piezoelectric Micromachined Ultrasonic Transducer with I-Shaped Composite Diaphragm. Micromachines. 2022;13:1597. doi: 10.3390/mi13101597. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 87.Wang H., Yu Y., Chen Z., Yang H., Jiang H., Xie H. Design and Fabrication of a Piezoelectric Micromachined Ultrasonic Transducer Array Based on Ceramic PZT; Proceedings of the 2018 IEEE Sensors; New Delhi, India. 28–31 October 2018; pp. 1–4. [Google Scholar]
  • 88.Cai J., Wang Y., Lou L., Zhang S., Gu Y., Gao F., Wu T. Photoacoustic and Ultrosound Dual-Modality Endoscopic Imaging Based on ALN Pmut Array; Proceedings of the 2022 IEEE 35th International Conference on Micro Electro Mechanical Systems Conference (MEMS); Tokyo, Japan. 9–13 January 2022; pp. 412–415. [Google Scholar]
  • 89.Zang J., Fan Z., Li P., Duan X., Wu C., Cui D., Xue C. Design and Fabrication of High-Frequency Piezoelectric Micromachined Ultrasonic Transducer Based on an AlN Thin Film. Micromachines. 2022;13:1317. doi: 10.3390/mi13081317. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 90.Cai J., Wang Y., Jiang D., Zhang S., Gu Y.A., Lou L., Gao F., Wu T. Beyond fundamental resonance mode: High-order multi-band ALN PMUT for in vivo photoacoustic imaging. Microsyst. Nanoeng. 2022;8:116. doi: 10.1038/s41378-022-00426-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 91.Shelton S., Chan M.-L., Park H., Horsley D., Boser B., Izyumin I., Przybyla R., Frey T., Judy M., Nunan K., et al. CMOS-compatible AlN piezoelectric micromachined ultrasonic transducers; Proceedings of the 2009 IEEE International Ultrasonics Symposium; Rome, Italy. 20–23 September 2009; pp. 402–405. [Google Scholar]
  • 92.Trolier-McKinstry S., Muralt P. Thin Film Piezoelectrics for MEMS. J. Electroceramics. 2004;12:7–17. doi: 10.1023/B:JECR.0000033998.72845.51. [DOI] [Google Scholar]
  • 93.Chen B., Chu F., Liu X., Li Y., Rong J., Jiang H. AlN-based piezoelectric micromachined ultrasonic transducer for photoacoustic imaging. Appl. Phys. Lett. 2013;103:031118. doi: 10.1063/1.4816085. [DOI] [Google Scholar]
  • 94.Yipeng L., Hao-Yen T., Fung S., Boser B.E., Horsley D.A. Short-range and high-resolution ultrasound imaging using an 8 MHz aluminum nitride PMUT array; Proceedings of the 2015 28th IEEE International Conference on Micro Electro Mechanical Systems (MEMS); Estoril, Portugal. 18–22 January 2015; pp. 140–143. [Google Scholar]
  • 95.Wasa K., Matsushima T., Adachi H., Kanno I., Kotera H. Thin-Film Piezoelectric Materials For a Better Energy Harvesting MEMS. J. Microelectromechanical Syst. 2012;21:451–457. doi: 10.1109/JMEMS.2011.2181156. [DOI] [Google Scholar]
  • 96.Luo G.-L., Kusano Y., Horsley D. Immersion PMUTs Fabricated with a Low Thermal-Budget Surface Micromachining Process; Proceedings of the 2018 IEEE International Ultrasonics Symposium (IUS); Kobe, Japan. 22–25 October 2018; pp. 1–4. [Google Scholar]
  • 97.Horsley D., Lu Y., Rozen O. Flexural Piezoelectric Resonators. In: Bhugra H., Piazza G., editors. Piezoelectric MEMS Resonators. Springer International Publishing; Cham, Swizterland: 2017. pp. 153–173. [Google Scholar]
  • 98.Muralt P. Which is the best thin film piezoelectric material?; Proceedings of the 2017 IEEE International Ultrasonics Symposium (IUS); Washington, DC, USA. 6–9 September 2017; pp. 1–3. [Google Scholar]
  • 99.Wang Q., Lu Y., Mishin S., Oshmyansky Y., Horsley D.A. Design, Fabrication, and Characterization of Scandium Aluminum Nitride-Based Piezoelectric Micromachined Ultrasonic Transducers. J. Microelectromechanical Syst. 2017;26:1132–1139. doi: 10.1109/JMEMS.2017.2712101. [DOI] [Google Scholar]
  • 100.Akiyama M., Umeda K., Honda A., Nagase T. Influence of scandium concentration on power generation figure of merit of scandium aluminum nitride thin films. Appl. Phys. Lett. 2013;102:021915. doi: 10.1063/1.4788728. [DOI] [Google Scholar]
  • 101.Kusano Y., Ishii I., Kamiya T., Teshigahara A., Luo G.L., Horsley D.A. High-SPL Air-Coupled Piezoelectric Micromachined Ultrasonic Transducers Based on 36% ScAlN Thin-Film. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2019;66:1488–1496. doi: 10.1109/TUFFC.2019.2921983. [DOI] [PubMed] [Google Scholar]
  • 102.Feng G.H., Liu H.J. Piezoelectric Micromachined Ultrasonic Transducers with a Cost-Effective Bottom-Up Fabrication Scheme for Millimeter-Scale Range Finding. Sensors. 2019;19:4696. doi: 10.3390/s19214696. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 103.Haynes W.M. CRC Handbook of Chemistry and Physics. 95th ed. CRC Press; Boca Raton, FL, USA: 2014. [DOI] [Google Scholar]
  • 104.Marioli D., Narduzzi C., Offelli C., Petri D., Sardini E., Taroni A. Digital time-of-flight measurement for ultrasonic sensors. IEEE Trans. Instrum. Meas. 1992;41:93–97. doi: 10.1109/19.126639. [DOI] [Google Scholar]
  • 105.Carotenuto R., Merenda M., Iero D., G. Della Corte F. Simulating Signal Aberration and Ranging Error for Ultrasonic Indoor Positioning. Sensors. 2020;20:3548. doi: 10.3390/s20123548. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 106.Koh Y., Wai Choong D.S., Chen D.S.-H., Goh D.J., Ghosh S., Sharma J., Merugu S., Quaglia F., Ferrera M., Savoia A.S., et al. High Resolution, High Frequency Ultrasonic Ranging in Air with pMUTs; Proceedings of the 2021 IEEE International Ultrasonics Symposium (IUS); Xi’an, China. 11–16 September 2021; pp. 1–4. [Google Scholar]
  • 107.Wu Z., Liu W., Tong Z., Zhang S., Gu Y., Wu G., Tovstopyat A., Sun C., Lou L. A Novel Transfer Function Based Ring-Down Suppression System for PMUTs. Sensors. 2021;21:6414. doi: 10.3390/s21196414. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 108.Pala S., Shao Z., Peng Y., Lin L. Improved Ring-Down Time and Axial Resolution of pMUTs via a Phase-Shift Excitation Scheme; Proceedings of the 2021 IEEE 34th International Conference on Micro Electro Mechanical Systems (MEMS); Gainesville, FL, USA. 25–29 January 2021; pp. 390–393. [Google Scholar]
  • 109.Huang Y.P., Wang J.S., Huang K.N., Ho C.T., Huang J.D., Young M.S. Envelope pulsed ultrasonic distance measurement system based upon amplitude modulation and phase modulation. Rev. Sci. Instrum. 2007;78:065103. doi: 10.1063/1.2745238. [DOI] [PubMed] [Google Scholar]
  • 110.Hong H., Yongtian W., Dayuan Y. A low-cost dynamic range-finding device based on amplitude-modulated continuous ultrasonic wave. IEEE Trans. Instrum. Meas. 2002;51:362–367. doi: 10.1109/19.997838. [DOI] [Google Scholar]
  • 111.Huang C.F., Young M.S., Li Y.C. Multiple-frequency continuous wave ultrasonic system for accurate distance measurement. Rev. Sci. Instrum. 1999;70:1452–1458. doi: 10.1063/1.1149604. [DOI] [Google Scholar]
  • 112.Chongchamsai M., Sinchai S., Wardkein P., Boonjun S. Distance Measurement Technique Using Phase Difference of Two Reflected Ultrasonic Frequencies; Proceedings of the 2018 3rd International Conference on Computer and Communication Systems (ICCCS); Nagoya, Japan. 27–30 April 2018; pp. 372–376. [Google Scholar]
  • 113.Dostart N., Brand M., Zhang B., Feldkhun D., Wagner K., Popović M.A. Vernier Si-Photonic Phased Array Transceiver for Grating Lobe Suppression and Extended Field-of-View; Proceedings of the Conference on Lasers and Electro-Optics; San Jose, CA, USA. 5–10 May 2019. [Google Scholar]
  • 114.Demi L. Practical Guide to Ultrasound Beam Forming: Beam Pattern and Image Reconstruction Analysis. Appl. Sci. 2018;8:1544. doi: 10.3390/app8091544. [DOI] [Google Scholar]
  • 115.Fain B., Blard F., Bastien J.-C., Gardien F., Frassati F. Beamforming with AlN-based bimorph piezoelectric micromachined ultrasonic transducers; Proceedings of the 2021 Smart Systems Integration (SSI); Grenoble, France. 27–29 April 2021; pp. 1–4. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Not applicable.


Articles from Micromachines are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES