Table 3.

Intra-day and inter-day average (avg) precision and accuracy assessed at four concentration levels (LLOQ, LQC, MQC and HQC) for five times (intra-day) in three different analytical runs (inter-day).

Intra-Day (n = 5) Inter-Day (n = 3)
QC Levels Nominal Conc.
(μg/mL)
Avg Conc. (μg/mL) Avg Precision (CV%) Avg Accuracy (Bias%) Avg Conc. (μg/mL) Avg Precision (CV%) Avg Accuracy (Bias%)
LLOQ 0.125 0.130 12.3 4.1 0.136 13.5 7.2
Low 7 6.55 10.2 9.1 6.41 7.6 9.3
Medium 70 68.3 8.4 5.8 64.9 7.0 7.9
High 280 252.4 11.6 9.9 241.7 13.0 13.7