Limits of detection (LODs) and limits of quantification (LOQs) for lyso-Gb3 and its analogues in dried blood spots (DBSs).

LOD LOQ
Whatman-GE 903 Capitainer®B Plasma Whatman-GE 903 Capitainer®B Plasma
nM nM nM nM nM nM
Lyso-Gb3-13C6 0.21 0.40 na 0.70 1.34 na
Lyso-Gb3 0.32 0.37 0.23 1.08 1.23 0.77
Lyso-Gb3 (Analogue − 28 Da) 0.11 0.18 0.06 0.36 0.59 0.21
Lyso-Gb3 (Analogue −2 Da) 0.22 0.15 0.29 0.73 0.55 0.97
Lyso-Gb3 (Analogue +16 Da) 0.10 0.39 0.22 0.32 1.29 0.72
Lyso-Gb3 (Analogue +18 Da) 0.13 0.23 0.14 0.42 0.77 0.47
Lyso-Gb3 (Analogue +34 Da) 0.13 0.19 0.24 0.43 0.64 0.79

LOD = limit of detection; LOQ = limit of quantification; na = not applicable.