Table 2.
Pilot In Vivo Insertion Data.
| Insertion |
|||||
|---|---|---|---|---|---|
| Animal: Side | Method | Depth, mm | Angle,º | Intracochlear trauma event | Electrocochleography notes |
| 1 | |||||
| Left | Device | 10.4 | 220 | EA touching BM, no elevation | 45% drop, signal recovered |
| Right | Manual | 10.0 | 235 | BM elevation | 60% drop, signal recovered and increased |
| 2 | |||||
| Left | Device | 9.9 | 201 | BM elevation | 50% drop, no signal recovery |
| Right | Manual | 9.4 | 219 | EA touching BM, no elevation | 55% drop, signal recovered and increased |
Abbreviations: BM, basilar membrane; EA, electrode array.