Table 2. Leads parameters of HBP and LBBP at the baseline and during the follow-up.
| 24 h after implantation | Post 6 months | Post 12 months | |
| Capture threshold, **P < 0.001 vs. HBP lead values. Sensing, &P < 0.001 vs. HBP lead values. Impendence, ##P < 0.01 and ###P < 0.001 vs. HBP lead values. HBP capture threshold increased after 6 months. HBP: His bundle pacing; LBBP: left bundle branch pacing. | |||
| HBP lead | |||
| Capture threshold, V | 0.95 ± 0.39 | 1.28 ± 0.83** | 1.39 ± 0.81** |
| Threshold increased to absolute value > 2 V, Number | 0 | 2 | 4 |
| Sensing | 3.94 ± 3.04 | 3.55 ± 3.16 | 3.43 ± 3.20 |
| Impendence, Ohm | 449.52 ± 77.60 | 441.00 ± 73.54 | 419.13 ± 75.74 |
| LBBP lead | |||
| Capture threshold | 0.68 ± 0.23** | 0.71 ± 0.22 | 0.70 ± 0.22 |
| Threshold increased to absolute value > 2 V, Number | 0 | 0 | 0 |
| Sensing | 12.01 ± 3.86& | 11.93 ± 3.88 | 12.07 ± 3.93 |
| Impendence, Ohm | 567.83 ± 168.89## | 579.57 ± 166.14### | 551.65 ± 135.11 |