Skip to main content
. 2022 Oct 29;106(2):401–411. doi: 10.1093/jaoacint/qsac130

Table 5.

Matrix study: 3M Petrifilm RYM Plate 25°C at 60 h versus DRBC—difference of means

Matrix Contamination level 3M Petrifilm RYM Plate, 25°C; 60 h
DRBC
DOMd SEe 90% CIa
95% CI
Meanb src Mean sr LCLf UCLg LCL UCL
Cannabis Flowerh Low 2.788 0.217 3.151 0.211 −0.363 0.113 −0.427 −0.299 −0.446 −0.280
High 5.392 0.166 5.539 0.143 −0.147 0.072 −0.302 0.007 −0.348 0.054
a

CI = Confidence interval for DOM.

b

Mean of five replicate portions, after logarithmic transformation: log10[cfu/g + (0.1)f] where f is the smallest reportable result.

c

Repeatability standard deviation.

d

DOM = Difference of means.

e

Standard error of the mean difference for paired analysis.

f

LCL = Lower confidence limit for DOM.

g

UCL = Upper confidence limit for DOM.

h

Matrix samples tested at Aurum Laboratories.