TABLE 1.

Validation parameters of meldonium’s biological method.

Parameter Value
Correlation coefficient of calibration curve r 2 = 0.9960
Range of calibration curve 500–50,000 ng/mL
Accuracy deviation −15.27%–1.18%
Intraseries precision 1.03%–5.57%
Interseries precision 4.63%–8.06%
Normalized matrix effect factor Coefficient of variation of matrix factor: 4.84% and 4.64%
Matrix selectivity (mean interference rate) 1.71%
Stability test at 4°C (relative deviation) 7 days: 1.19% ∼ 3.45%; 21 days: 2.25% ∼ 2.27%
Stability test with freeze–thaw at −20°C 7 days: 91.6%–95.19%; 21 days: 81.8%–89.01%
Stability test at room temperature 4 h: 94.32%–96.9%
Carryover effect ≤20% of LLOQ value
Rate of recovery 99.07% ∼ 105.28%