Skip to main content
. 2023 Feb 23;10:1080101. doi: 10.3389/fcvm.2023.1080101

Table 3.

Summary of orthotopic and heterotopic tricuspid valve replacement device characteristics.

EVOQUE NaviGate LuX valve TriSol Intrepid TricValve Tricento
Manufacturer Edwards Lifesciences NaviGate Cardiac Structures Jenscare Bio-technology TriSol Medical Medtronic Products + Features New Valve Technology
Frame and Design Nitinol frame with fabric skirt, nine anchors Cone-shaped nitinol stent with 12 tines and 12 atrial winglets Nitinol stent with atrial disc, interventricular septal anchor, two graspers Cone-shaped nitinol frame with six fixation arms Dual-stent self-expanding nitinol Two self-expanding nitinol stents Self-expanding nitinol
Pericardial Leaflets Bovine Equine Bovine Bovine Bovine Bovine Porcine
Components 1 1 1 1 1 2 1
Anchoring TV leaflets/ annulus TV leaflets/ annulus Septal anchor and anterior leaflet grasp Tricuspid annulus Perimeter oversizing Stent expansion/ radial force with device oversizing Stent expansion/ radial force with device oversizing
Radial Force Dependent Dependent Independent Dependent Dependent Dependent Dependent
Sizes (mm) 44,48,52 36,40,44, 48,52 50,60,70 (annulus) 62.5 (outflow) 50.3 (inflow) 43,46,50 Up to 38 (SVC) Up to 43 (IVC) Up to 48
Delivery Femoral Jugular / Atrial Atrial / Mini-thoracotomy Jugular Femoral / Apical Femoral Femoral
Recapturable No No No Yes until fixations arms fully expanded Yes Yes (to 80% deployed) No
Delivery system size (Fr) 28 42 32 30 35 27.5 24

Fr, French size; SVC, superior vena cava; IVC, inferior vena cava.