Nominal DUV (ng mL−1) Mean DUV (ng mL−1) Recovery (%) RSD (%)
Intra-day
5 4.76 95.12 1.15
15 15.33 102.21 2.15
150 146.19 97.46 0.70
400 398.37 99.59 0.24
Inter-day
5 4.75 94.95 2.69
15 15.13 100.88 1.21
150 146.47 97.64 1.38
400 398.23 99.56 0.75