Supporting Figure 6

Fig. 6.

Fluorescent titration of Fluo-4 with Ca2+. Shown is the percent of maximal fluorescence versus the calculated [Ca2+] (μM). The titration produced the expected noncooperative binding curve and a Fluo-4 Kd for Ca2+ of 450 ± 50 nM was determined, a value close to the literature Kd of 365 nM. The agreement between the Kd of Fluo-4 calculated from this experiment with its literature value confirms that the Ca2+/EGTA buffer is accurately calibrated to give the proper [Ca2+].