Table 2.

Bland-Altman values and percentage errors for each stepwise change to the image acquisition and analysis technique.

MEAN VALUES
MAXIMUM VALUES
Overall mean Difference
95% limits of agreement % error Overall mean Difference
95% limits of agreement % error
Mean Std. Dev Mean Std. Dev
ORIGINAL APPROACH
       Whole Valve, ungated PET, non-contrast CT
ORIGINAL Standard Uptake Value 1.523 0.008 0.194 -0.373 to 0.389 26 1.955 0.094 0.263 -0.421 to 0.608 27
ORIGINAL Tissue to Background Ratio (using brachiocephalic) 1.439 0.047 0.451 -0.836 to 0.930 63 1.869 0.154 0.612 -1.045 to 1.354 65
        RA blood pool correction
corrected Standard Uptake Value (subtracting RA) 0.425 0.008 0.091 -0.170 to 0.186 43 0.858 0.094 0.168 -0.236 to 0.423 39
Tissue to Background Ratio (using RA) 1.418 -0.002 0.086 -0.171 to 0.167 12 1.842 0.064 0.200 -0.328 to 0.456 22
        Most Diseased Segment Approach
Standard Uptake Value 1.652 0.007 0.207 -0.400 to 0.413 25 2.129 0.131 0.265 -0.389 to 0.651 25
Tissue to Background Ratio (using RA) 1.547 -0.005 0.075 -0.151 to 0.142 10 2.012 0.107 0.145 -0.177 to 0.391 14

FINAL APPROACH
       RA blood pool, most diseased segment, gated PET, contrast CT
FINAL Standard Uptake Value 1.662 0.044 0.291 -0.527 to 0.615 35 2.528 0.275 0.633 -0.966 to 1.515 50
FINAL Tissue to Background Ratio (using RA) 1.546 -0.046 0.078 -0.199 to 0.107 10 2.385 0.111 0.439 -0.750 to 0.971 37