Skip to main content
. Author manuscript; available in PMC: 2017 Mar 22.
Published in final edited form as: Phys Med Biol. 2016 Mar 17;61(7):2851–2878. doi: 10.1088/0031-9155/61/7/2851

Table 6.

Results of the signal-to-background (S2B) evaluation of the image reconstructions of the rabbit-sized phantom.

Trig Normalized S2B (with TOF) S2B (no TOF) TOF change
Un-tilted 3 Decays 2.41 2.25 7.10%
1 All 2.69 2.23 20.33%
3 Counts 2.37 2.21 6.85%
3 All 2.41 2.25 7.15%

Tilted 3 All 2.31 2.15 7.68%
1 Decays 2.59 2.13 21.82%
3 Counts 2.31 2.15 7.53%
1 All 2.61 2.14 22.14%

Note: the signal-to-background values are reported for each of the reconstructions shown in figures 11 and 12. In the last column, the relative change of the signal-to-background value using the TOF information compared to the non-TOF case is reported.