Table 1.

Acquisition parameters of the T1w and T2w sequences. Patient habitus determined acquisition parameters. Using a lower bandwidth of 630 Hz for the T1w sequence (patients 7 to 10) resulted in slightly improved results of the joint MoCo-HDTV reconstruction. Fat suppression was used in 8 out of 10 cases for the T2w sequence.

Parameter T1w T2w
Number of patients 10 10
Orientation Axial Axial
Echo train length N.A. 62
Number of slices 88-96 80-96
Slice resolution (%) 100.0 80.0-100.0
Slice oversampling (%) 16.7-27.7 25.0-40.0
Number of spokes per slice 935-1440 N.A.
Total acquisition time (min) 04:58-07:42 05:34-09:16
Field of view (mm3) 336x336x288-384x384x290 420x242x264-480x288x290
Voxel size (mm3) 1.31x1.31x3.0-1.5x1.5x3.3 1.31x1.31x3.0-1.5x1.5x3.3
Echo time (ms) 1.24-1.55 108-115
Repetition time (ms) 2.51-3.18 1000-1500
Flip angle (°) 8-12 90
Pixel bandwidth (Hz) 630-1085 560-600
Fat suppression Yes 8:Yes, 2:No