Published in final edited form as: NMR Biomed. 2019 Mar 29;32(6):e4095. doi: 10.1002/nbm.4095

Table 2.

Reproducibility values of human cardiac 31P-MRS from the literature

Reference Field strength (T) Cohort size Acquisition time (min) Localisation method Nominal voxel size (cm3) Absolute reproducibility (%)a
This study 7 10 28 3D-CSI 1.5×1.5×2.5 = 5.6mL 26
3D-CSI 2.5×1.5×2.5 = 9.4mL 39

Bakermans et al, 201724 3 8 7 Single voxel 3D ISIS (8.0)3 = 512mL 40
1D ISIS with 1D CSI - 43
1D ISIS with 1D CSI - 44

Tyler et al, 200915 3 20 30 3D CSI 1.5×1.5×2.5 = 5.6mL 53

Shivu et al, 201029 3 1b 23 3D ISIS 4.4×5.5×3.7 = 90mL 31

Schaefer et al, 199230 1.5 7 7 1D CSI - 22

Lamb et al, 199625 1.5 16 10 3D ISIS 6.0×7.0×7.0 = 294mL 38 c
15 1D CSI - -
30 2D ISIS + 1D CSI 1.0cm thick slices 46 c
a

Absolute reproducibility is calculated as (CR ÷ Mean PCr/ATP) × 100%

b

A single subject was scanned 8 times to assess reproducibility.

c

Absolute reproducibility was computed by extracting raw values from Figure 5 in ref. 25 using WebPlotDigitizer (https://apps.automeris.io/wpd/)