Skip to main content
. Author manuscript; available in PMC: 2019 Oct 3.
Published in final edited form as: Med Phys. 2019 Jul 9;46(9):3767–3775. doi: 10.1002/mp.13586

Table I.

Hypofractionated PMRT dose objectives for target coverage and organs-at-risk

Structure Dose metric Preferred objective Acceptable limit

Targets Volume >95% Rx >95% n/a
Targets Volume >90% Rx n/a >90%
Heart Mean dose <4 Gy <6 Gy
Volume >25 Gy <7% <10%
Ipsilateral lung Volume >17 Gy <35% <40%

Rx, prescription; n/a, not applicable.