Skip to main content
. Author manuscript; available in PMC: 2020 Aug 6.
Published in final edited form as: Magn Reson Med. 2020 Apr 14;84(5):2364–2375. doi: 10.1002/mrm.28278

Table 3:

Myocardium and blood SNR of Cartesian, EPI and spiral in-out bSSFP cine acquisitions using 0.55T. SNR is measured at end-diastole and peak systole in a mid-ventricular short axis slice.

Cartesian bSSFP EPI bSSFP Spiral in-out bSSFP
Myocardium SNR Diastole 4.9 ± 0.9 7.0 ± 1.5 8.4 ± 2.4
Systole 5.8 ± 1.0 8.5 ± 2.1 10.6 ± 2.3
Blood SNR Diastole 13.8 ± 3.0 17.3 ± 4.9 22.7 ± 5.7
Systole 13.9 ± 3.3 18.5 ± 5.2 22.8 ± 5.2
Blood-myocardium CNR Diastole 8.9 ± 2.2 10.3 ± 3.6 14.3 ± 3.8
Systole 8.0 ± 2.3 10.0 ± 3.4 12.2 ± 3.1