Skip to main content
. Author manuscript; available in PMC: 2022 Oct 14.
Published in final edited form as: Magn Reson Med. 2022 Jun 17;88(5):2043–2057. doi: 10.1002/mrm.29352

Table 1.

Quantitative values of T1, T2 and ADC from mdMRF-LTE and –STE scans from four volunteers.

N = 4 WM- Frontal WM- Parietal GM - Putamen
T1 ref (ms) 900.54 ± 24.86 937.55 ± 46.6 1241.90 ±57.74
mdMRF-LTE 879.87 ± 28.57 917.76 ±33.08 1187.92 ±35.15
mdMRF-STE 887.27 ± 25.04 921.42 ±21.11 1195.14 ±38.37
T2 ref (ms) 36.67 ±1.78 41.29 ±3.30 44.51 ±3.93
mdMRF-LTE 47.57 ± 0.95 53.26 ±2.41 49.92 ±1.42
mdMRF-STE 47.64 ± 0.16 53.31 ±1.43 49.15 ± 0.80
ADC ref (μm2/ms) 0.77 ± 0.01 0.83 ± 0.04 0.71 ± 0.01
mdMRF-LTE 0.75 ± 0.06 0.69 ± 0.07 0.56 ± 0.02
mdMRF-STE 0.76 ± 0.16 0.79 ± 0.17 0.66 ± 0.09