Skip to main content
. Author manuscript; available in PMC: 2011 Jan 1.
Published in final edited form as: J Appl Clin Med Phys. 2010 Apr 19;11(2):3185. doi: 10.1120/jacmp.v11i2.3185

Fig. 12.

Fig. 12

MapCHECK analysis of a TPS calculation that imperfectly models the high gradients of a DMLC IMRT field. Failing points are captured, but the lack of detector density does not make the source of error obvious.