Published in final edited form as: J Appl Clin Med Phys. 2010 Apr 19;11(2):3185. doi: 10.1120/jacmp.v11i2.3185

Table IV.

Percent difference/DTA comparison between EPID converted dose planes to 2D diode array measurements for seven designed test fields (Varian linac) and two analysis criteria (% difference/DTA): 3%/3mm and 2%/2mm (DTA). The number of distinct diode measurement positions in the fields is shown, as some of these fields were quite small and did not cover many diodes in the MapCHECK array.

Field Description 3mm/3% pass rate 2mm/2% pass rate Points Analyzed
1.0 × 1.0 MLC, 10 × 10 Primary Jaws 100.0 100.0 6
1.5 × 2.0 MLC, 10 × 10 Primary Jaws 100.0 91.7 12
2.0 × 1.0 MLC, 10 × 10 Primary Jaws 100.0 93.3 15
3.0 × 3.0 MLC, 10 × 10 Primary Jaws 100.0 100.0 28
5.0 × 5.0 MLC, 10 × 10 Primary Jaws 97.1 91.3 69
7.0 × 7.0 MLC, 10 × 10 Primary Jaws 94.2 88.3 121
Inverted Dose Pyramid
(11 MLC segments,
fixed 10×10 Primary Jaws)
99.2 88.4 250