Published in final edited form as: Magn Reson Med. 2013 Nov 18;72(4):1162–1169. doi: 10.1002/mrm.25015

Table 2.

Acquisition parameters of PC-MRI sequences.

Machine 3 Tesla MR system (Trio, Siemens AG, Erlangen, Germany)
Coil standard 32-channel phased-array coil
Sequence k-space segmented 3D RF-spoiled gradient-echo sequence with interleaved three-directional velocity encoding
Gating Respiratory navigator gating and prospective ECG gating
Spatial resolution 1.25–1.77 × 1.25–1.77 × 3.2mm3
Matrix size 192–256 × 120–192
Field of view Rectangular: 320–340 to 200–256) mm2
Number of slices 20–40
Temporal resolution 40ms
Velocity encoding 1.5 m/s
Frames per cardiac cycle 16–20
Flip angle
TR 5ms
TE 2.519ms
Segmentation factor 2