Published in final edited form as: Radiat Prot Dosimetry. 2009 Dec 3;138(4):10.1093/rpd/ncp278. doi: 10.1093/rpd/ncp278

Table 3.

The differences in dose values between experiment and simulation with and without breast bismuth on the body phantom (mGy).

Position kVp Body phantom without bismuth Body phantom with bismuth
Mea Sim Differ(%) Mea Sim Differ(%)
Center 120 0.95 0.97 −2.28 0.81 0.83 −2.75
80 0.27 0.26 5.14 0.22 0.21 4.22
12h 120 1.75 1.70 2.68 1.03 1.01 1.81
80 0.62 0.58 6.45 0.28 0.27 3.74
3h 120 1.75 1.71 2.67 1.66 1.64 1.33
80 0.62 0.58 6.12 0.58 0.56 2.99
6h 120 1.51 1.54 −1.95 1.50 1.53 −2.14
80 0.51 0.52 −0.16 0.50 0.51 −3.40
9h 120 1.77 1.73 2.39 1.68 1.68 0.36
80 0.63 0.59 7.52 0.62 0.57 8.94
1

Mea: Measurement, Sim: Monte Carlo simulation, Differ: Difference between measurement and simulation