Table 1.

Technical specifications of the SACRTD

X-ray source: GE Seifert X-ray tube Isovolt titan 225 M2
Model: NDI 225M metal ceramic dual focal spot
Focal spot size: 0.4 mm/3 mm
Power: 640 W/3000 W
Inherent filtration 0.8 mm Be
Additional filtration: 0.5 mm Cu
Generator: GE IsoVolt Titan 225
Target angle: 20°
Robot manufacturer: Adept Inc, Livermore, CA ,USA Model: Adept Viper s650
Overall arm length: 565mm (270 mm first arm and 295 second arm)
Motion Range: J1±1700, J2=−1900, +45, J3=−290,+2560,J4=±1900, J5=±1200, J6=±3600
Maximum Pay load: 5 kg
Positional repeatability: ±0.02 mm in each of X, Y and Z direction,
Drive motor and break: AC servo motors for all joints and breaks for joint J2 andJ6
Controller: Adept Smart controller and Adept T1 pendent
Flat-panel x-ray detector: Perkin-Elmer, Wiesbaden, Germany
Model: XRD 0820 CN3 digital x-ray detector
Detector size: 20.48 cm×20.48 cm
Pixel matrix format: 1024×1024
Pixel pitch: 0.2 mm
Scintillator: CsI (Tl)
Maximum frame rate: 7 fps