Published in final edited form as: Tomography. 2016 Dec;2(4):411–420. doi: 10.18383/j.tom.2016.00229

Table 1.

Acquisition Parameters for tDWI and dDWI Pulse Sequences in Diffusion Phantom and In Vivo

Phantom In Vivo (Abdomen) In Vivo (Pelvis)
tDWI dDWI tDWI dDWI tDWI dDWI
Orientation Coronal Coronal Axial Axial Axial oblique Axial oblique
Diffusion scheme Monopolar Monopolar Monopolar Monopolar Monopolar Monopolar
Gradient directions [x,y,z] [1.0, 1.0, −0.5], [1.0, −0.5, 1.0], [−0.5, 1.0, 1.0] [1.0, 1.0, 1.0] [1.0, 1.0, −0.5], [1.0, −0.5, 1.0], [−0.5, 1.0, 1.0] [1.0 1.0 1.0] [1.0, 1.0, −0.5], [1.0, −0.5, 1.0], [−0.5, 1.0, 1.0] [1.0, 1.0, 1.0]
Read-out SS SE-EPI SS SE-EPI SS SE-EPI SS SE-EPI SS SE-EPI SS SE-EPI
b Values (s/mm2) 0, 500, 900, 2000 0, 500, 900, 2000 50, 400, 800 50, 400, 800 50, 1000, 1600 50, 1000, 1600
Averages 4, 4, 4, 4 4, 12, 12, 12 1, 2, 4 3, 6, 12 1, 5, 10 3, 15, 30
TE (ms) 110 (1.5 T); 99 (3.0 T) 106 (1.5 T); 95 (3.0 T) 73 (1.5 T); 70 (3.0 T) 70 (1.5 T); 68 (3.0 T) 83 (1.5 T); 71 (3.0 T) 79 (1.5 T); 69 (3.0 T)
TR (ms) 10 000 10 000 5000 5000 5000 5000
Slice thickness (mm) 5 5 5 5 5 5
Number of slices 5 5 30 30 20 20
Matrix 128 × 128 128 × 128 128 × 128 128 × 128 114 × 114 114 × 114
FOV (mm2) 210 × 210 210 × 210 370 × 250 370 × 250 250 × 250 250 × 250
GRAPPA acceleration 2 2 2 2 2 2
Fat suppression None None SPAIR SPAIR SPAIR SPAIR
Acquisition time (min) 7:10 7:10 2:12 2:12 4:27 4:27

The in vivo acquisitions were performed during free breathing.

Abbreviations: dDWI, diagonal diffusion-weighted imaging; FOV, field of view; GRAPPA, generalized autocalibrating partial parallel acquisition; SPAIR, spectral attenuated inversion recovery; tDWI, 3-scan trace diffusion-weighted imaging; SS SE-EPI, single-shot spin-echo echo planar imaging; TE, echo time; TR, repetition time.