Published in final edited form as: J Appl Clin Med Phys. 2017 Jun 2;18(4):84–96. doi: 10.1002/acm2.12100

Table 6.

comparison between ultrasound monitored movement and movement tracked by camera. All numbers are in mm. LR: left and right direction, AP: Anterior and posterior direction, SI: superior and Inferior direction.

US Tracking Results Camera tracked (ground truth) Difference
Run ID LR AP SI LR AP SI LR AP SI
1 2.30 1.50 9.20 2.37 1.31 8.87 0.07 −0.19 −0.33
2 2.10 1.30 9.20 2.34 1.25 8.79 0.24 −0.05 −0.41
3 2.20 1.50 9.20 2.33 1.36 8.85 0.13 −0.14 −0.35
4 2.30 1.40 9.10 2.31 1.33 8.86 0.01 −0.07 −0.24
5 2.20 1.30 9.10 2.34 1.27 8.85 0.14 −0.03 −0.25
6 2.10 1.20 9.10 2.35 1.27 8.83 0.25 0.07 −0.27
7 2.20 1.20 9.00 2.35 1.24 8.84 0.15 0.04 −0.16
8 2.20 1.20 9.10 2.37 1.26 8.90 0.17 0.06 −0.20
9 2.20 1.20 9.00 2.35 1.28 8.87 0.15 0.08 −0.13
10 2.30 1.40 9.00 2.34 1.24 8.86 0.04 −0.16 −0.14
mean 2.21 1.32 9.10 2.35 1.28 8.85 0.14 −0.04 −0.25
SD 0.07 0.12 0.08 0.02 0.04 0.03 0.08 0.10 0.09