Table 1. 3T brain MRI anatomical acquisition protocols.

Each of the 7 sites used one of 3 different Siemens scanner models (Skyra, Verio and TrioTim), necessitating 3 model-specific protocols for each of the 3 pulse sequences.

3D T2 FLAIR 3D T1 MPRAGE 2D PD T2
Scanner manufacturer Siemens Skyra Siemens Verio Siemens TrioTim Siemens Skyra Siemens Verio Siemens TrioTim Siemens Skyra Siemens Verio Siemens TrioTim
Operation system version Syngo MR D13 Syngo MR B17 Syngo MR B17 Syngo MR D13 Syngo MR B17 Syngo MR B17 Syngo MR D13 Syngo MR B17 Syngo MR B17
Coil 20, 32, or 64 Channel* 32 Channel 32 Channel 20, 32, or 64 Channel* 32 Channel 32 Channel 20, 32, or 64 Channel* 32 Channel 32 Channel
Acceleration factor for parallel imaging 2 2 2 2 2 2 3 3 3
Orientation Sagittal Sagittal Sagittal Sagittal Sagittal Sagittal Axial Axial Axial
Field of view (cm) 25.6×25.6 25.6×25.6 25.6×25.6 25.6×25.6 25.0×25.0 25.0×25.0 21.7×24.0 21.7×24.0 21.7×24.0
Matrix size 512×512 512×512 512×512 256×256 256×256 256×256 232×256 232×256 232×256
Number of slices 176 176 176 176 176 176 108 108 108
Repetition time (msec) 4800 4800 4800 1900 1900 1900 3000 3000 3000
Echo time (msec) 353 354 355 2.52 2.52 2.52 11/101 11/101 11/101
Flip angle (degrees) 120 120 120 9 9 9 150 150 150
Voxel size (mm) 0.5×0.5×1.0 0.5×0.5×1.0 0.5×0.5×1.0 1.0×1.0×1.0 0.977×0.977×1.0 0.977×0.977×1.0 0.9375×0.9375×3.0 0.9375×0.9375×3.0 0.9375×0.9375×3.0
Scan time (mins) 6:53 7:00 7:00 4:15 4:16 4:16 1:57 2:15 2:15
Number of signal averages 1 1 1 1 1 1 1 1 1

Key: FLAIR = fluid-attenuated inversion recovery, MPRAGE = magnetization-prepared rapid acquisition gradient echo, PD = proton density.

*

Brigham and Women’s Hospital = 20-channel, University of California San Francisco = 64 channel (the other two Skyra sites = 32 channel).